Johnson A P, Dowdall T, Brown M C, Hammond P E
J Med Eng Technol. 1985 Sep-Oct;9(5):218-26. doi: 10.3109/03091908509032102.
Equipment management arrangements in hospitals, including safety and function testing, relating to continuous-flow anaesthetic apparatus have been found wanting following examination of a number of recent incidents. A problem exists with the complexity of the tests required and the number of necessary test instruments. In this article the required tests are examined, and a specially designed test set to make fuller testing a practical proposition for hospital staff and for manufacturers' service technicians is described.