Analytical evaluation of a direct ion-selective-based analyser: Still gaps to close.

机构信息

Department of Laboratory Medicine, Ghent University Hospital, Ghent, Belgium.

Department of Laboratory Medicine, Ghent University Hospital, Ghent, Belgium.

出版信息

Clin Biochem. 2024 Dec;133-134:110829. doi: 10.1016/j.clinbiochem.2024.110829. Epub 2024 Sep 28.

Abstract

INTRODUCTION

Discrepancies between electrolyte concentrations determined by blood gas analysers (BGA) and core-lab chemistry analysers may create confusion in clinical practice. This problem is rooted in the different ion-selective electrode (ISE) methodologies that are used. Whilst most available chemistry analysers use indirect ISE, we evaluated the analytical performance of the new automated chemistry analyser Biossays™ E6 (Snibe), equipped with direct ISE, for the determination of sodium (Na), potassium (K), chloride (Cl), ionized calcium (iCa) and pH.

MATERIALS AND METHODS

Total precision, estimated deviation and total error were evaluated for all analytes on the E6 analyser. Several patient cohorts were used to perform method comparisons between the E6 and the direct (RP500e BGA) and indirect (Architect c16000 analyser) ISE methods routinely used in the lab. Obtained data were compared against pre-set quality specifications and used for adjustment of the 2 direct ISE methods. For Na and iCa, agreement with the routinely used protein-corrected Na and total calcium (TCa) concentrations were assessed respectively.

RESULTS

The analytical performance for the 4 tested electrolytes (Na, K, Cl, iCa) and pH were acceptable and within the specified performance specifications. After adjustment of both direct methods, method comparison on an independent patient cohort showed good agreement. For Na and iCa, a good correlation with the protein corrected Na and TCa results was observed.

CONCLUSION

The acceptable analytical performance and ease-of-use of the E6 direct ion selective instrument is making it feasible to optimize electrolyte determinations to direct methodology.

摘要

简介

血气分析仪(BGA)和核心实验室化学分析仪测定的电解质浓度之间存在差异,这可能会给临床实践带来困惑。这个问题源于所使用的不同离子选择性电极(ISE)方法学。虽然大多数可用的化学分析仪使用间接 ISE,但我们评估了新的自动化化学分析仪 Biossays™ E6(Snibe)的分析性能,该仪器配备了直接 ISE,用于测定钠(Na)、钾(K)、氯(Cl)、离子化钙(iCa)和 pH。

材料与方法

我们评估了 E6 分析仪上所有分析物的总精密度、估计偏差和总误差。使用几个患者队列,在实验室常规使用的直接(RP500e BGA)和间接(Architect c16000 分析仪)ISE 方法之间进行 E6 与方法比较。将获得的数据与预设的质量规格进行比较,并用于调整 2 种直接 ISE 方法。对于 Na 和 iCa,分别评估了与常规使用的蛋白校正 Na 和总钙(TCa)浓度的一致性。

结果

4 种测试电解质(Na、K、Cl、iCa)和 pH 的分析性能可接受,且符合规定的性能规格。在调整了两种直接方法后,在独立的患者队列上进行方法比较显示出良好的一致性。对于 Na 和 iCa,与蛋白校正的 Na 和 TCa 结果有很好的相关性。

结论

E6 直接离子选择性仪器的可接受的分析性能和易用性使其能够优化电解质的直接方法学测定。