Urology Unit, ASST Fatebenefratelli-Sacco, Milano, Italy.
Urology Department, Polytechnic University of Le Marche, Ancona, Italy.
World J Urol. 2024 Oct 3;42(1):560. doi: 10.1007/s00345-024-05275-9.
A new digital single-use flexible ureteroscope, Pusen direct in scope suction (DISS) 7.5Fr (PU3033AH), was evaluated with respect to manoeuvrability, suction quality, visibility and clinical efficiency.
A prospective cohort study was conducted in six tertiary reference centers in Europe and Asia between February-April 2024. Adult patients who underwent flexible ureteroscopy and laser lithotripsy (fURSL) for urolithiasis were included. Demographic, intraoperative and follow-up characteristics were recorded. Quality parameters were rated by each surgeon using a Likert scale.
a total of 57 fURSL were performed. Preoperative characteristics revealed a mean stone volume of 480.00mm (mean Hounsfield Unit- 998). 57.9%(n = 33) of the patients were pre-stented, and a ureteric access sheath was used in 64.9%(n = 37). Integrated-suction was deemed helpful in 94.7%(n = 54) fURSL as reported by the operators. An initial stone free rate (SFR) confirmed by postoperative imaging was achieved in 84.21%(n = 48). Quality parameters of the scope reached a mean Likert score of 4.5, with a "very good" mean evaluation for "scope placement", "visual quality", "irrigation", "deflection", "manoeuvrability", and "weight". "Suction quality" and "overall performance satisfaction" were rated "good". Comparing the Pusen DISS scope with other previously used scopes, overall satisfaction was rated 4.1. When asked if the surgeons would be willing to use the Pusen DISS 7.5Fr scope in the future, all the six surgeons answered positively.
The DISS 7.5 Pusen ureteroscope displayed good visibility, manoeuvrability and suction quality, with excellent operative results. Further evaluation with larger comparative cohorts will help understand the potential of in-vivo use of integrated suctioning systems for fURSL.
评估一种新型数字一次性使用柔性输尿管镜,Pusen 直接镜内吸引(DISS)7.5Fr(PU3033AH),在操控性、吸引质量、可视性和临床效率方面的性能。
2024 年 2 月至 4 月,在欧洲和亚洲的 6 个三级参考中心进行了一项前瞻性队列研究。纳入接受软性输尿管镜激光碎石术(fURSL)治疗尿路结石的成年患者。记录患者的人口统计学、术中及随访特征。每位外科医生使用李克特量表对质量参数进行评分。
共进行了 57 例 fURSL。术前特征显示结石平均体积为 480.00mm(平均 Hounsfield 单位为 998)。57.9%(n=33)的患者术前放置了支架,64.9%(n=37)使用了输尿管接入鞘。操作人员报告,集成吸引在 94.7%(n=54)的 fURSL 中被认为是有帮助的。术后影像学证实初始结石清除率(SFR)为 84.21%(n=48)。该输尿管镜的质量参数评分为 4.5,“镜体放置”、“视觉质量”、“冲洗”、“弯曲”、“操控性”和“重量”的平均评分为“非常好”。“吸引质量”和“整体性能满意度”评为“良好”。与其他先前使用的输尿管镜相比,总体满意度评分为 4.1。当被问及外科医生是否愿意在未来使用 Pusen DISS 7.5Fr 输尿管镜时,6 位外科医生均给予肯定回答。
DISS 7.5 Pusen 输尿管镜具有良好的可视性、操控性和吸引质量,手术效果极佳。进一步使用更大的队列进行评估将有助于了解体内使用集成吸引系统进行 fURSL 的潜力。