The purpose of this study was to assess local diagnostic reference levels (LDRLs) for full-field digital mammography (FFDM) and digital breast tomosynthesis (DBT) mammography in India. Data from 1500 women were collected from five different mammography facilities in major cities in Tamil Nadu, India. The mean of mean glandular dose were used to arrive at an LDRL. The noted mean compressed breast thickness was 55.26 ± 3.4. The recorded mean MGDs for the five centres were 3.1 ± 0.1 and 3.8 ± 0.2 mGy for FFDM and DBT, respectively. The 75th percentile value for all five centers is 3.3 and 4.0 mGy for FFDM and DBT, respectively. The LDRLs found in the current study were also compared with those from earlier studies conducted in other nations, such as the United Kingdom, Malaysia, Morocco, and Ghana. The present study is the first of its kind to determine the LDRL for the FFDM and DBT scanners operating in the Tamil Nadu region, India, and is proposed as a starting point that will allow professionals to evaluate and optimize their practice. Furthermore, similar studies in other regions of India are necessary in order to establish National DRLs.
本研究旨在评估印度全数字化乳腺摄影术(FFDM)和数字乳腺断层合成术(DBT)的局部诊断参考水平(LDRL)。从印度泰米尔纳德邦五个主要城市的五个不同的乳腺摄影设施中收集了 1500 名女性的数据。使用平均腺体剂量的平均值来确定 LDRL。记录的平均压缩乳房厚度为 55.26 ± 3.4。五个中心的记录平均 MGD 分别为 FFDM 和 DBT 的 3.1 ± 0.1 和 3.8 ± 0.2 mGy。所有五个中心的第 75 个百分位数值分别为 3.3 和 4.0 mGy。本研究还将当前研究中的 LDRL 与英国、马来西亚、摩洛哥和加纳等其他国家进行的早期研究进行了比较。本研究是首次确定印度泰米尔纳德邦地区运行的 FFDM 和 DBT 扫描仪的 LDRL,并建议作为一个起点,使专业人员能够评估和优化他们的实践。此外,印度其他地区的类似研究对于建立国家 DRL 是必要的。