Faculty of Sciences, Moulay Ismail University, Meknès, Morocco.
Faculty of Sciences,Ibn Tofail University, Kenitra, Morocco.
J Med Imaging Radiat Sci. 2022 Jun;53(2):242-247. doi: 10.1016/j.jmir.2022.03.008. Epub 2022 Mar 29.
The objective of this study was to establish local diagnostic reference levels (LDRLs) for the full-field digital mammography (FFDM) and tomosynthesis (DBT) in Moroccan health facilities.
Data from 146 women were collected from three facilities. The proposed DRLs were defined as the 75th percentile of the mean average glandular dose (AGD) distribution.
The mean AGD recorded in this study for the three centers was 1.47 mGy for all centers, and 1.42 mGy and 1.64 mGy for the CC and MLO projections, respectively. The mean compressed breast thickness (CBT) values recorded in this current study were 55 mm, the LDRLs reported for all centers was 1.7 mGy, the CC projection was 1.6 mGy, and the MLO projection was 1.8 mGy. In addition, the LDRLs reported in the current study were compared with those from previous studies for other countries, including the United Kingdom, Japan, Ghana, and Sri Lanka.
This work provides an assessment of local DRLs for mammography in Morocco and is suggested as a starting point that will allow professionals to evaluate and optimize their practice. Furthermore, the definition of national DRLs is a necessary process in optimizing Moroccan medical exposures.
本研究旨在为摩洛哥医疗机构的全数字化乳腺摄影(FFDM)和断层合成(DBT)建立局部诊断参考水平(LDRL)。
从三个中心收集了 146 名女性的数据。所提出的 LDRL 被定义为平均腺体剂量(AGD)分布的第 75 个百分位数。
本研究中三个中心的平均 AGD 记录为所有中心的 1.47 mGy,CC 和 MLO 投影的平均 AGD 分别为 1.42 mGy 和 1.64 mGy。本研究中记录的平均压缩乳房厚度(CBT)值为 55 mm,报告的所有中心 LDRL 为 1.7 mGy,CC 投影为 1.6 mGy,MLO 投影为 1.8 mGy。此外,本研究报告的 LDRL 与包括英国、日本、加纳和斯里兰卡在内的其他国家的先前研究报告的 LDRL 进行了比较。
本工作对摩洛哥的乳腺摄影进行了局部 DRL 评估,被建议作为一个起点,以允许专业人员评估和优化他们的实践。此外,国家 DRL 的定义是优化摩洛哥医疗照射的必要过程。