Galloway S J, Rossiter G J, Pidcock M E
Pathology. 1985 Oct;17(4):563-7. doi: 10.3109/00313028509084753.
The Coulter Diff3 50 analyzes Wright stained films using computerized pattern recognition. It is capable of reporting on the differential leukocyte count, red cell morphology, platelet sufficiency and the reticulocyte count. Although the precision of the differential leukocyte and reticulocyte counts is better than that of the manual method, correlation is poor and monocyte classification is a major problem. The Diff3 50 is useful for examining relatively normal films and for follow-up counts, but in our experience it has proven more unreliable and costly than expected.
库尔特Diff3 50利用计算机模式识别技术分析瑞氏染色涂片。它能够报告白细胞分类计数、红细胞形态、血小板充足情况以及网织红细胞计数。虽然白细胞分类计数和网织红细胞计数的精度优于手工方法,但相关性较差,单核细胞分类是一个主要问题。Diff3 50对于检查相对正常的涂片和进行后续计数很有用,但根据我们的经验,它已被证明比预期更不可靠且成本更高。