Department of Perio-Prosthodontic Implantology, Graduate School of Clinical Dentistry, Ajou University, Suwon, Republic of Korea.
Department of Prosthodontics, Institute of Oral Health Science, Ajou University School of Medicine, Suwon, Republic of Korea.
Sci Rep. 2024 Oct 9;14(1):23640. doi: 10.1038/s41598-024-74354-z.
This study aimed to explore the clinical applicability of the shade-matching function in intraoral scanners. This study measured the tooth colors of maxillary anterior dentitions of 83 adults using visual matching, a spectrophotometer, and a scanner according to two color systems: VITA Classical (VC) and VITA 3D-Master (V3D). Agreement between each method was assessed by weighted Cohen's kappa coefficient (K, α = 0.05). For V3D, the overall agreement between the scanner and spectrophotometer (K = 0.498) was higher than that between the scanner and visual matching (K = 0.473). Similarly, the agreement between the scanner and spectrophotometer (K = 0.283) was higher than that between a scanner and visual matching (K = 0.140) for VC. Regarding tooth position, the highest agreement between the scanner and spectrophotometer was observed on the right central incisor (K = 0.542) for V3D. Tooth color measurement with a scanner was comparable to that with a spectrophotometer, especially on the central incisors when using the VITA 3D-Master system. A scanner could serve as an alternative to a spectrophotometer for shade selection. However, color matching should still be visually verified.
本研究旨在探讨口内扫描仪在比色方面的临床适用性。本研究采用视觉比色、分光光度计和扫描仪,按照两个比色系统(VITA Classical [VC] 和 VITA 3D-Master [V3D])对 83 名成年人上颌前牙的牙色进行了测量。采用加权 Cohen's kappa 系数(K,α=0.05)评估每种方法之间的一致性。对于 V3D,扫描仪与分光光度计之间的总体一致性(K=0.498)高于扫描仪与视觉比色之间的一致性(K=0.473)。同样,对于 VC,扫描仪与分光光度计之间的一致性(K=0.283)高于扫描仪与视觉比色之间的一致性(K=0.140)。就牙位而言,V3D 中右中切牙的扫描仪与分光光度计之间的一致性最高(K=0.542)。与分光光度计相比,使用 VITA 3D-Master 系统时,扫描仪测量牙色的结果相当,尤其是在中切牙上。扫描仪可以作为分光光度计的替代工具用于比色选择。然而,颜色匹配仍应进行视觉验证。