The effect of patient shielding on fetal radiation dose was evaluated in computed tomography pulmonary angiography with the out-of-plane shield visible in the localizer but absent in the scan range in chest computed tomography (CT). An anthropomorphic phantom with additional prosthetic pregnancy belly was scanned with different CT scanners using clinical imaging protocols and radiophotoluminescence dosemeters (GD-352 M). The out-of-plane shield decreased the fetal absorbed radiation dose with Siemens Somatom go.Up, Canon Aquilion Prime SP and Canon Aquilion One scanners. The decrease was 3.9%-39.4% (0.01-0.09 mGy). With GE Optima the shield increased the fetal dose by 100% (0.23 mGy), with Canon Aquilion One and GE Optima scanners the abdomen dose increase was 17.5% and 36.4%, respectively (0.61 and 1.38 mGy). Applying an out-of-plane shield outside the scanned volume may increase the fetal radiation dose during CT when using tube current modulation, depending on the make and model of the CT scanner.
在平面外屏蔽可见于定位像但在胸部 CT 扫描范围内缺失的情况下,评估了平面外屏蔽对胎儿辐射剂量的影响。使用不同的 CT 扫描仪和放射性荧光剂量计(GD-352 M)对具有附加假体妊娠腹部的人体模型进行了扫描。西门子 Somatom go.Up、佳能 Aquilion Prime SP 和佳能 Aquilion One 扫描仪的平面外屏蔽降低了胎儿吸收的辐射剂量。降低幅度为 3.9%-39.4%(0.01-0.09 mGy)。GE Optima 屏蔽使胎儿剂量增加 100%(0.23 mGy),佳能 Aquilion One 和 GE Optima 扫描仪的腹部剂量增加分别为 17.5%和 36.4%(0.61 和 1.38 mGy)。当使用管电流调制时,在 CT 期间,在扫描体积外应用平面外屏蔽可能会增加胎儿的辐射剂量,具体取决于 CT 扫描仪的制造商和型号。