J Prosthodont. 2025 Apr;34(4):382-387. doi: 10.1111/jopr.13974. Epub 2024 Oct 30.
This in vitro study aimed to compare the 3-dimensional (3D) accuracy of intraoral and desktop scanners when digitizing the post and core polyvinyl siloxane impressions.
Ten extracted human teeth were prepared to build a post space in the root canal. Each tooth was scanned using a micro-computed tomography device (MCT), and the slice data were reconstructed as controls. A conventional quadrant polyvinyl siloxane impression was made and scanned thrice using an intraoral scanner (IOS) and a desktop blue-light scanner (IMS), with one of the scans randomly selected as the definitive cast. Precision was assessed by comparing the deviations among repeated scans. Trueness was evaluated by analyzing the deviations between the MCT and each definitive cast (MCT-IOS and MCT-IMS). The clinically acceptable root mean square (RMS) value was set to 50 µm.
The mean RMS values of the repeated scans of IOS and IMS were 18.3 ± 3.9 µm and 13.9 ± 6.2 µm (P < 0.05), and the difference between them was 4.4 ± 5.9 µm. The mean RMS values of MCT-IOS and MCT-IMS were 28.3 ±3.0 µm and 31.3 ± 4.2 µm (P < 0.05), with a difference of 3.0 ± 2.9 µm. The RMS values were significantly lower than 50 µm (P < 0.05).
The IOS showed slightly better trueness than the IMS but slightly lower precision. The precision and trueness of the two scanners for digitizing post and core impressions were acceptable for clinical application.
本体外研究旨在比较口内扫描仪和台式扫描仪在对桩核聚乙烯基硅氧烷印模进行数字化处理时的三维(3D)精度。
制备10颗拔除的人牙,在根管内构建桩道。使用微型计算机断层扫描设备(MCT)对每颗牙齿进行扫描,并将切片数据重建作为对照。制作传统的象限聚乙烯基硅氧烷印模,使用口内扫描仪(IOS)和台式蓝光扫描仪(IMS)对其进行三次扫描,其中一次扫描随机选为最终模型。通过比较重复扫描之间的偏差来评估精度。通过分析MCT与每个最终模型(MCT - IOS和MCT - IMS)之间的偏差来评估准确性。临床上可接受的均方根(RMS)值设定为50µm。
IOS和IMS重复扫描的平均RMS值分别为18.3±3.9µm和13.9±6.2µm(P<0.05),两者差值为4.4±5.9µm。MCT - IOS和MCT - IMS的平均RMS值分别为28.3±3.0µm和31.3±4.2µm(P<0.05),差值为3.0±2.9µm。RMS值显著低于50µm(P<0.05)。
IOS在准确性方面略优于IMS,但精度略低。两种扫描仪对桩核印模进行数字化处理的精度和准确性在临床应用中是可以接受的。