Patients with head and neck cancers often require radiotherapy, where immobilization devices like thermoplastic masks ensure precise radiation delivery by minimizing movement. However, the quality of these masks lacks standard reference data. This study aimed to establish institutional acceptance criteria for thermoplastic mask quality and quantify their effectiveness using a 3 dimensional stress test and verified the setup errors using daily megavoltage computed tomography (MVCT). Between April and June 2022, 30 patients underwent radiotherapy with thermoplastic masks. Four key facial points (forehead, bilateral cheekbones, and chin) were tested for supporting force. Mean forces ranged from 3.97 N to 8.8 N. MVCT was used to assess 6 dimensional-axis errors, with mean translational errors (x, y, z) of 0.32 mm, -1.09 mm, and 2.24 mm, respectively, and rotational errors (yaw, pitch, roll) of -0.12°, 0.22°, and 0.35°, respectively. The results demonstrated that the thermoplastic masks provided precise immobilization, minimizing setup errors in 6 dimensions. Our findings offer a quantifiable method to ensure high-quality immobilization during radiotherapy for patients with head and neck cancers.
头颈部癌症患者常需接受放射治疗,热塑面罩等固定装置通过最大限度减少移动来确保精确的放射治疗。然而,这些面罩的质量缺乏标准参考数据。本研究旨在建立热塑面罩质量的机构可接受标准,并使用三维压力测试来量化其效果,并使用每日兆伏 CT(MVCT)验证设置误差。2022 年 4 月至 6 月期间,30 名患者接受了热塑面罩放射治疗。对四个关键面部点(前额、双侧颧骨和下巴)进行支撑力测试。平均力范围为 3.97N 至 8.8N。MVCT 用于评估 6 个轴的误差,平均平移误差(x、y、z)分别为 0.32mm、-1.09mm 和 2.24mm,旋转误差(偏航、俯仰、滚动)分别为-0.12°、0.22°和 0.35°。结果表明,热塑面罩提供了精确的固定,最大限度地减少了 6 个维度的设置误差。我们的研究结果提供了一种可量化的方法,以确保头颈部癌症患者放射治疗期间的高质量固定。