Repeatability of 3D MR fingerprinting during scanner software upgrades.

作者信息

Dupuis Andrew, Chen Yong, Chow Kelvin, Griswold Mark A, Boyacioglu Rasim

机构信息

Department of Biomedical Engineering, Case Western Reserve University, Cleveland, OH, USA.

Department of Radiology, Case Western Reserve University, Cleveland, OH, USA.

出版信息

MAGMA. 2024 Nov 5. doi: 10.1007/s10334-024-01211-5.

Abstract

OBJECTIVE

This study aims to quantify the repeatability of a 3D Magnetic Resonance Fingerprinting (MRF) research protocol in the context of a scanner software upgrade. All of MRI assumes consistent hardware performance and raw data pre-processing on the acquisition side. Software upgrades can affect hardware specifications and reconstruction chain parameters. Understanding how vendor-provided software upgrades vary MRF-derived T1 and T2 values is crucial for its application in different settings.

MATERIALS AND METHODS

Eight healthy volunteers were imaged with an in-house developed 3D MRF pulse sequence using a 3T scanner before and after a software upgrade (VA31A to VA50A, MAGNETOM Vida, Siemens Healthineers). Online MRF reconstruction using Singular Value Decomposition (SVD) timeseries compression and B1+ correction was performed. The study involved test-retest repeatability assessment and a comparison of pre- and post-upgrade data based on automatically extracted T1 and T2 values from MNI-152 Harvard-Oxford Subcortical Structural Atlas regions.

RESULTS

Significant mismatches were found directly after the upgrade. However, after an information exchange with the vendor, the 3D-MRF sequence showed consistent repeatability in both intra-version test-retest scenarios and cross-version comparisons: 1.16 ± 3.18% variability in T1 and 0.54 ± 4.84% in T2 for intra-version tests, and 0.83 ± 3.68% (T1) and 0.05 ± 5.81% (T2) variability for cross-version comparisons.

DISCUSSION

The study shows the reliable performance of 3D MRF protocols across software upgrades is possible, but it also highlights the importance of detailed evaluation and vendor collaboration in ensuring consistency. These findings support the application of MRF in longitudinal studies and emphasize the need for systematic assessments following hardware or software modifications.

摘要

目的

本研究旨在量化在扫描仪软件升级的情况下,三维磁共振指纹识别(MRF)研究方案的可重复性。所有磁共振成像都假定采集端的硬件性能和原始数据预处理是一致的。软件升级可能会影响硬件规格和重建链参数。了解供应商提供的软件升级如何改变基于MRF得出的T1和T2值,对于其在不同环境中的应用至关重要。

材料与方法

8名健康志愿者在软件升级(从VA31A升级到VA50A,MAGNETOM Vida,西门子医疗)前后,使用3T扫描仪通过内部开发的三维MRF脉冲序列进行成像。使用奇异值分解(SVD)时间序列压缩和B1 +校正进行在线MRF重建。该研究包括重测可重复性评估,以及基于从MNI - 152哈佛 - 牛津皮质下结构图谱区域自动提取的T1和T2值,对升级前后的数据进行比较。

结果

升级后立即发现了显著的不匹配。然而,在与供应商进行信息交流后,三维MRF序列在版本内重测场景和跨版本比较中均显示出一致的可重复性:版本内测试中,T1的变异性为1.16±3.18%,T2的变异性为0.54±4.84%;跨版本比较中,T1的变异性为0.83±3.68%,T2的变异性为0.05±5.81%。

讨论

该研究表明,跨越软件升级的三维MRF方案具有可靠的性能,但也强调了在确保一致性方面进行详细评估和与供应商合作的重要性。这些发现支持了MRF在纵向研究中的应用,并强调了在硬件或软件修改后进行系统评估的必要性。