Medical Radiation Physics and Nuclear Medicine, Karolinska University Hospital, Stockholm, Sweden.
Medical Radiation Physics and Nuclear Medicine, Karolinska University Hospital, Stockholm, Sweden; Department of Clinical Science, Intervention and Technology, Karolinska Institutet, Sweden.
Phys Med. 2024 Aug;124:103426. doi: 10.1016/j.ejmp.2024.103426. Epub 2024 Jul 9.
To analyze the image quality of a novel, state-of-the art platform for CBCT image-guided spine surgery, focusing particularly on the dose-effectiveness compared with conventional CT (the gold standard for postoperative assessment).
The ClarifEye platform (Philips Healthcare) with integrated augmented-reality surgical navigation, has been compared with a GE Revolution CT (GE Healthcare). The 3D spatial resolution (TTF) and noise (NPS) were evaluated considering relevant feature contrasts (200-900 HU) and background noise for differently sized patients (200-300 mm water-equivalent diameter). These measures were used to determine the noise equivalent quanta (NEQ) and observer model detectability.
The CBCT system exhibited a linear response with 50% TTF at 5.7 cycles/cm (10% TTF at 9.2 cycles/cm), and the axial noise power peaking at about 3.6 cycles/cm (average frequency of 4.1 cycles/cm). The noise magnitude and texture differed markedly compared to iteratively reconstructed CT images (GE ASiR-V). The CBCT system had 26% lower detectability for a high-frequency task (related to edge detection) compared with CT images reconstructed using the Bone kernel combined with ASiR-V 50%. Likewise, it had 18% lower detectability for low- and mid-frequency tasks compared with CT images reconstructed using the Standard kernel. This difference translates to 50%-80% higher CBCT imaging doses required to match the CT image quality.
The ClarifEye platform demonstrates intraoperative CBCT-imaging capabilities that under certain circumstances are comparable with conventional CT. However, due to limited dose-effectiveness, a trade-off between timeliness and radiation exposure must be considered if end-of-procedure CBCT is to replace postoperative CT.
分析一种新型的、最先进的 CBCT 图像引导脊柱手术平台的图像质量,特别关注其与传统 CT(术后评估的金标准)的剂量-效果比较。
采用集成增强现实手术导航的 ClarifEye 平台(飞利浦医疗保健公司)与 GE Revolution CT(通用电气医疗保健公司)进行比较。评估了三维空间分辨率(TTF)和噪声(NPS),考虑了相关特征对比度(200-900 HU)和不同大小患者的背景噪声(200-300 mm 水当量直径)。这些措施用于确定噪声等效量子(NEQ)和观察者模型可探测性。
CBCT 系统表现出线性响应,50% TTF 时为 5.7 个周期/cm(10% TTF 时为 9.2 个周期/cm),轴向噪声功率峰值约为 3.6 个周期/cm(平均频率为 4.1 个周期/cm)。与迭代重建 CT 图像(GE ASiR-V)相比,噪声幅度和纹理差异显著。CBCT 系统对于高频任务(与边缘检测相关)的可探测性比使用 Bone 内核结合 ASiR-V 50%重建的 CT 图像低 26%。同样,与使用 Standard 内核重建的 CT 图像相比,低频和中频任务的可探测性低 18%。这意味着要匹配 CT 图像质量,需要将 CBCT 成像剂量提高 50%-80%。
ClarifEye 平台展示了术中 CBCT 成像能力,在某些情况下可与传统 CT 相媲美。然而,由于剂量效果有限,如果要在手术结束时用 CBCT 代替术后 CT,就必须在及时性和辐射暴露之间进行权衡。