Nakahara T, Yoshida F
J Biomed Mater Res. 1986 Mar;20(3):363-74. doi: 10.1002/jbm.820200308.
Empirical studies were conducted on how rates of hemolysis were influenced by mechanical factors, such as shear, turbulence, agitation, pressure fluctuation, and bubbling, in five types of experimental apparatus: rotating cylinder, rotating disks, agitated vessel, pressure pulsator, and bubble columns. Our data indicate, at least qualitatively, that an important mechanical factor controlling rates of hemolysis in practical devices, in which blood does not flow in thin channels or capillaries, is the turbulent stress in the bulk blood, rather than the shear at the blood-solid interface.
在五种实验装置(旋转圆筒、旋转圆盘、搅拌容器、压力脉动器和鼓泡塔)中,针对诸如剪切力、湍流、搅拌、压力波动和气泡形成等机械因素如何影响溶血速率开展了实证研究。我们的数据至少定性地表明,在血液并非在细通道或毛细血管中流动的实际装置中,控制溶血速率的一个重要机械因素是血液主体中的湍流应力,而非血液与固体界面处的剪切力。