两种扫频源光学相干断层扫描生物测量仪测量黄斑脱离孔源性视网膜脱离眼眼轴长度的准确性

Accuracy of Axial Length Measurements by Two Swept-Source Optical Coherence Tomography Biometers in Macula-off Rhegmatogenous Retinal Detachment Eyes.

作者信息

Helaly Hany Ahmed, Elnaggar Osama Ramadan, Abou Shousha Mohsen, Elhady Amr Mohamed

机构信息

Ophthalmology Department, Faculty of Medicine, Alexandria University, Alexandria, Egypt.

出版信息

Clin Ophthalmol. 2024 Nov 19;18:3321-3334. doi: 10.2147/OPTH.S469094. eCollection 2024.

Abstract

PURPOSE

To assess the accuracy of axial length (AXL) measurements using two swept-source optical coherence biometers, IOLMaster 700 (Carl Zeiss Meditec AG, Jena, Germany) and ARGOS (Alcon, Inc. Fort Worth, TX), for macula-off rhegmatogenous retinal detachment (RRD).

METHODS

This retrospective study included 100 eyes with phakic primary macula-off RRD. Preoperative AXL measurements were performed using different methods: applanation A-scan ultrasound (U/S) biometry, combined applanation vector A/B-scan biometry, and optical biometry measurements were obtained using IOLMaster 700 (Carl Zeiss, Meditec, Jena, Germany) and ARGOS (Alcon, Inc. Fort Worth, TX). All patients underwent pars plana phacovitrectomy. At 8-10 weeks postoperatively, optical biometry was performed to record AXL.

RESULTS

Mean preoperative AXL measured using vector-A/B-scan ultrasonography was higher than that of postoperative AXL measured using IOLMaster (p < 0.05). Mean AXL measured by the standard mode of the ARGOS optical biometer was lower than mean AXL measured by both enhanced retina visualization (ERV) mode and user-adjusted method (p < 0.05). Mean same-eye AXL measured using IOLMaster was lower than that measured using ARGOS (p < 0.05). The least difference was observed with combined vector-A/B-scan ultrasound (on the positive side), followed by fellow eye AXL measured using IOLMaster optical biometry (on the negative side).

CONCLUSION

Optical biometry of fellow eye in macula-off RRD was noted to be highly correlating with postoperative optical biometry of same eye using IOL Master 700 in eyes without anisometropia. IOLMaster 700 showed less accuracy in the AXL measurements for same eye. The ARGOS optical biometer may have a good potential for measuring same eye AXL. Using ERV mode or a user-adjusted method for the ARGOS optical biometer may improve accuracy of AXL measurements. Most accurate method for measuring AXL in same eye was vector-A/B-scan ultrasound.

摘要

目的

评估使用两种扫频光学相干生物测量仪(IOLMaster 700,德国耶拿卡尔蔡司医疗技术公司;ARGOS,美国德克萨斯州沃思堡爱尔康公司)测量黄斑脱离性孔源性视网膜脱离(RRD)患者眼轴长度(AXL)的准确性。

方法

这项回顾性研究纳入了100只患有晶状体眼原发性黄斑脱离性RRD的眼睛。术前使用不同方法测量AXL:压平A超生物测量法、联合压平矢量A/B超生物测量法,以及使用IOLMaster 700(德国耶拿卡尔蔡司医疗技术公司)和ARGOS(美国德克萨斯州沃思堡爱尔康公司)进行的光学生物测量。所有患者均接受了睫状体平坦部晶状体玻璃体切除术。术后8 - 10周,进行光学生物测量以记录AXL。

结果

使用矢量A/B超超声测量的术前平均AXL高于使用IOLMaster测量的术后AXL(p < 0.05)。ARGOS光学生物测量仪标准模式测量的平均AXL低于增强视网膜可视化(ERV)模式和用户调整方法测量的平均AXL(p < 0.05)。使用IOLMaster测量的同眼平均AXL低于使用ARGOS测量的结果(p < 0.05)。联合矢量A/B超超声测量的差异最小(在正值方面),其次是使用IOLMaster光学生物测量法测量的对侧眼AXL(在负值方面)。

结论

在无屈光参差的眼中,黄斑脱离性RRD患者对侧眼的光学生物测量结果与使用IOL Master 700测量的同眼术后光学生物测量结果高度相关。IOLMaster 700在测量同眼光轴长度时准确性较低。ARGOS光学生物测量仪在测量同眼光轴长度方面可能具有良好的潜力。对ARGOS光学生物测量仪使用ERV模式或用户调整方法可能会提高光轴长度测量的准确性。测量同眼光轴长度最准确的方法是矢量A/B超超声。