应用扫频源光学相干生物测量仪增强视网膜可视化模式测量致密性白内障眼轴长度。

Evaluation of Axial Length Measurement Using Enhanced Retina Visualization Mode of the Swept-Source Optical Coherence Tomography Biometer in Dense Cataract.

机构信息

Department of Ophthalmology, Japan Community Healthcare Organization Chukyo Hospital, Nagoya, Japan.

Department of Ophthalmology, Keio University School of Medicine, Tokyo, Japan.

出版信息

Ophthalmic Res. 2021;64(4):595-603. doi: 10.1159/000515054. Epub 2021 Feb 5.

Abstract

INTRODUCTION

It has been reported that even using the swept-source (SS) optical coherence biometer, it is challenging to measure the axial length (AL) in cases with advanced cataracts. The enhanced retina visualization (ERV) mode, which is equipped with OCTB1 (ARGOS), shifts the peak of measurement sensitivity to the retinal side so that the AL can be measured even if the light energy is attenuated. The aim of the present study was to evaluate the accuracy and efficacy of the ERV mode in measuring the AL of dense cataracts.

METHODS

This was a single-center retrospective observational case series conducted in Japan. We included 213 eyes of 213 consecutive patients with advanced cataracts who underwent preoperative evaluation. The AL was measured before and after surgery using two SS optical coherence tomography biometers (OCTB1 and OCTB2; IOLMaster 700). Cases in which OCTB1 the standard mode failed to measure AL, OCTB1 with the ERV mode was used instead. Primary outcome measures were the acquisition rate and the AL measurement accuracy using the ERV mode. The χ2 test, the Kruskal-Wallis test, and the Wilcoxon signed-rank test were used to compare the acquisition rate and differences between pre- and postoperative AL values, respectively. In the ERV subgroup, Bland-Altman plots were used to compare the pre- and postoperative AL values measured using OCTB1-ERV mode. A p-value of less than 5% was considered statistically significant.

RESULTS

The AL acquisition rate was not significantly different between OCTB1 with the standard mode and OCTB2. The AL of 65 eyes (30.5%) could not be measured using OCTB1 with the standard mode. Conversely, the AL of 51 of these eyes (78.5%) was successfully measured using OCTB1 with the ERV mode. In these 51 eyes, a difference of ≤0.2 mm and of ≤0.1 mm between pre- and postoperative AL measurements was observed in 40 (78.4%) and 30 eyes (58.8%), respectively. The Bland-Altman plot found no systematic error between pre- and postoperative AL values measured using the ERV mode.

CONCLUSION

In patients with dense cataracts, AL measurement using the standard mode of an SS-OCT biometer is challenging. Furthermore, the ERV mode could be promising for AL measurement in such cases.

摘要

简介

据报道,即使使用扫频源(SS)光相干生物测量仪,在白内障严重的情况下测量眼轴(AL)也具有挑战性。增强型视网膜可视化(ERV)模式配备了 OCTB1(ARGOS),将测量灵敏度的峰值移向视网膜侧,因此即使光能衰减,也可以测量 AL。本研究旨在评估 ERV 模式测量致密白内障 AL 的准确性和有效性。

方法

这是一项在日本进行的单中心回顾性观察性病例系列研究。我们纳入了 213 例 213 例连续接受术前评估的严重白内障患者的 213 只眼。使用两台 SS 光相干断层扫描仪(OCTB1 和 OCTB2;IOLMaster 700)分别在术前和术后测量 AL。当 OCTB1 标准模式无法测量 AL 时,改用 OCTB1 的 ERV 模式。主要观察指标为 ERV 模式的采集率和 AL 测量精度。采用卡方检验、Kruskal-Wallis 检验和 Wilcoxon 符号秩检验分别比较采集率和术前与术后 AL 值的差异。在 ERV 亚组中,采用 Bland-Altman 图比较 OCTB1-ERV 模式测量的术前和术后 AL 值。p 值<0.05 为差异有统计学意义。

结果

OCTB1 标准模式与 OCTB2 的 AL 采集率无显著差异。65 只眼(30.5%)的 AL 无法用 OCTB1 标准模式测量。相反,这 65 只眼中的 51 只(78.5%)的 AL 可以用 OCTB1 的 ERV 模式成功测量。在这 51 只眼中,40 只(78.4%)和 30 只(58.8%)的术后 AL 测量值与术前差值分别≤0.2mm 和≤0.1mm。ERV 模式测量的术前和术后 AL 值的 Bland-Altman 图未发现系统误差。

结论

在白内障严重的患者中,SS-OCT 生物测量仪的标准模式测量 AL 具有挑战性。此外,ERV 模式在这种情况下测量 AL 可能很有前途。