Multack Sam, Plummer Nellie, Smits Gerard, Hall Brad
Multack Eye Care, Frankfort, IL, USA.
CSC, Inc., Santa Barbara, CA, USA.
Clin Ophthalmol. 2023 Aug 17;17:2423-2428. doi: 10.2147/OPTH.S407538. eCollection 2023.
To compare the prediction accuracy of the Argos biometer using standard keratometry to the prediction accuracy of the IOLMaster 700 biometer using Total Keratometry.
This was a randomized, prospective, single surgeon study of 80 right eyes of 80 patients that had preoperative biometry with both the Argos and IOLMaster 700 devices, followed by cataract surgery and intraocular lens (IOL) implantation. Prediction errors (directional and absolute) for each device were determined from the 1 month postoperative manifest refraction.
The directional prediction error was 0.07 ± 0.32 D for the Argos and 0.08 ± 0.34 D for the IOLMaster 700. The mean of the difference in prediction error (directional) was 0.02 D, which was not statistically significant (p > 0.05). The absolute prediction error was 0.21 ± 0.25 D for the Argos and 0.25 ± 0.24 D for the IOLMaster 700. The mean of the difference in absolute prediction error was 0.04 D, which was statistically significant (p < 0.004) but not clinically significant. The percentage of eyes with absolute prediction error ≤ 0.5 D was 91% (73 eyes) for the Argos and 88% (70 eyes) for the IOLMaster 700. This difference was not statistically significant.
The prediction accuracies were similar between the Argos and IOLMaster 700 in eyes with normal axial length. There was a significant difference in mean absolute prediction error between devices; however, this was not clinically meaningful.
比较使用标准角膜曲率测量法的阿高斯生物测量仪与使用全角膜曲率测量法的IOLMaster 700生物测量仪的预测准确性。
这是一项随机、前瞻性、单术者研究,纳入80例患者的80只右眼,术前使用阿高斯和IOLMaster 700设备进行生物测量,随后进行白内障手术和人工晶状体(IOL)植入。根据术后1个月的明显验光结果确定每种设备的预测误差(方向和绝对误差)。
阿高斯的方向预测误差为0.07±0.32D,IOLMaster 700为0.08±0.34D。预测误差(方向)差异的平均值为0.02D,无统计学意义(p>0.05)。阿高斯的绝对预测误差为0.21±0.25D,IOLMaster 700为0.25±0.24D。绝对预测误差差异的平均值为0.04D,具有统计学意义(p<0.004),但无临床意义。绝对预测误差≤0.5D的眼睛百分比,阿高斯为91%(73只眼),IOLMaster 700为88%(70只眼)。这种差异无统计学意义。
在眼轴长度正常的眼中,阿高斯和IOLMaster 700的预测准确性相似。两种设备之间的平均绝对预测误差存在显著差异;然而,这在临床上并无意义。