Heterophile antibodies, which can arise from infections, autoimmune disorders, or exposure to animal antigens, can interfere with immunoassays. These antibodies can cross-react with the test reagents used in troponin assays, causing a false elevation in troponin levels. The paper describes a case of a 37-year-old male drug abuser admitted to the emergency room with chest pain. A series of troponin measurements performed using different assays gave discrepant results. Only thanks to the use of Scantibodies HBT tubes, which remove heterophile antibodies, was it possible to make a correct diagnosis of troponin negativity. In conclusion, a correct laboratory/clinical approach to the identification of heterophile antibody interference is essential for accurate troponin testing in order to avoid false positive results. Implementing neutralizing tests can significantly improve the reliability of these diagnostic assays, ensuring better patient outcome.
嗜异性抗体可由感染、自身免疫性疾病或接触动物抗原引起,可能会干扰免疫测定。这些抗体可能与肌钙蛋白检测中使用的检测试剂发生交叉反应,导致肌钙蛋白水平假性升高。本文描述了一例 37 岁男性药物滥用者,因胸痛被收入急诊室。使用不同检测方法进行的一系列肌钙蛋白测量结果不一致。只有使用 Scantibodies HBT 管(可去除嗜异性抗体)才能正确诊断为肌钙蛋白阴性。总之,正确的实验室/临床方法来识别嗜异性抗体干扰对于准确进行肌钙蛋白检测至关重要,以避免假阳性结果。实施中和检测可以显著提高这些诊断检测的可靠性,确保更好的患者预后。