Thoröe-Boveleth Sven, Becker Ruth, Bertram Jens, Schettgen Thomas, Möller Manfred, Jonigk Danny, Kraus Thomas, Weiskirchen Ralf
Institute for Occupational, Social and Environmental Medicine (IASU), Medical Faculty, RWTH University Hospital Aachen, Pauwelsstraße 30, 52074, Aachen, Germany.
Institute of Pathology, Medical Faculty, RWTH University Hospital Aachen, Pauwelsstraße 30, 52074, Aachen, Germany.
Anal Sci. 2025 Mar;41(3):237-249. doi: 10.1007/s44211-024-00691-8. Epub 2024 Nov 27.
Calibrations in LA-ICP-MS are typically very time-consuming and complex, as they need to be matched to the samples being measured and sectioned on a microtome. Alternatively, gelatin can be in droplet form or as a section, which is a more recent development. In this study, we report on investigations where hot multi-element gelatin solutions are placed in a linear fashion on microscopic slides to conduct comparative statistical observations between doped tissue homogenates from the liver and lung. The tissue homogenates served as both samples (complete ablation) and calibration standards (partial ablation) for verification purposes. We explored the effects of different laser fluences (0.50-1.50 J/cm), gelatin contents (0.3-20.0%) and section thicknesses (10-30 µm). To do this, we evaluated the samples by calculating median and mean values over the entire section with and without removal of elementary spikes (de-spiking). A reduction in distribution was achieved with averaging. The data was normalized using C as an internal standard. In these experiments and under these measurement conditions, it was observed that the selected laser fluences, gelatin contents, and section thicknesses did not visibly affect the results, making them comparable. Each sample could be assessed with each gelatin calibration, allowing for determination of expected reference values. Despite interruptions in the measurement operation, due to the high number of measurements, where samples and calibrations could not be analyzed in one measurement run, no negative effects of stopping and starting the LA-ICP-MS were observed.
激光烧蚀电感耦合等离子体质谱(LA-ICP-MS)中的校准通常非常耗时且复杂,因为它们需要与在切片机上测量和切片的样品相匹配。另外,明胶可以呈液滴形式或制成切片,这是一项较新的进展。在本研究中,我们报告了相关研究,即将热的多元素明胶溶液以线性方式放置在载玻片上,以便对来自肝脏和肺的掺杂组织匀浆进行比较统计观察。出于验证目的,组织匀浆既用作样品(完全烧蚀),也用作校准标准品(部分烧蚀)。我们探究了不同激光能量密度(0.50 - 1.50 J/cm)、明胶含量(0.3 - 20.0%)和切片厚度(10 - 30 µm)的影响。为此,我们通过计算在去除和未去除元素尖峰(去尖峰)情况下整个切片的中位数和平均值来评估样品。通过平均实现了分布的减少。数据以碳作为内标进行归一化。在这些实验以及这些测量条件下,观察到所选的激光能量密度、明胶含量和切片厚度并未对结果产生明显影响,从而使它们具有可比性。每个样品都可以用每种明胶校准进行评估,从而能够确定预期的参考值。尽管测量操作中存在中断情况,由于测量次数众多,样品和校准无法在一次测量运行中进行分析,但未观察到停止和启动LA-ICP-MS产生负面影响。