An assessment of seven closed system transfer devices in accordance with the 2015 NIOSH vapor containment performance protocol.

作者信息

Armistead Lori T, Stepanovic Michael, Earnhart Megan A, Eckel Stephen F

机构信息

Division of Practice Advancement and Clinical Education, University of North Carolina at Chapel Hill Eshelman School of Pharmacy, Chapel Hill, NC, USA.

出版信息

J Oncol Pharm Pract. 2024 Dec 10:10781552241304638. doi: 10.1177/10781552241304638.

Abstract

INTRODUCTION

The purpose of this study was to assess and compare the effectiveness of seven different closed system transfer device (CSTD) product lines following the 2015 NIOSH Vapor Containment Performance Protocol, using a Gasmet™ DX5000 Terra multigas FTIR analyzer.

METHODS

Seven different CSTD systems were assessed using a two-task evaluation on their capacity to contain the NIOSH-specified challenge agent, 70% isopropyl alcohol (IPA). Task 1 simulated reconstitution and compounding steps while Task 2 simulated compounding and administration steps. Vial adapters, syringe adapters, IV bag adapters, and y-site adapters from each CSTD product line were tested. All tasks were performed in a custom-built environmental test chamber as outlined in the 2015 NIOSH protocol, with isopropyl alcohol (IPA) release detected and measured by a Gasmet™ analyzer.

RESULTS

The BD PhaSeal™ CSTD product line effectively contained IPA vapor during both tasks with a task performance metric of 0.00 ppm for each task. BD PhaSeal™ Optima, Equashield, Yukon Medical Arisure, ICU Medical ChemoLock™, BD Texium™, and ICU Medical ChemoClave™ non-vented CSTDs demonstrated detectable and quantifiable IPA leakage during each task. The leaks occurred at different steps within the protocol, and the amount of IPA vapor detected varied.

CONCLUSION

In this study, one CSTD product line (i.e., PhaSeal™) successfully contained IPA vapor below the Gasmet™ analyzer's limit of quantification (LOQ) for IPA of 0.04 ppm for both Tasks 1 and 2, demonstrating it to be a validated closed system per the study protocol. All other CSTD product lines demonstrated some level of IPA leakage with task performance metrics above the analyzer's LOQ. However, these results tell us little about how well CSTDs contain hazardous drugs (HDs), as a universal HD surrogate for effectively assessing the closedness of all types of CSTDs has yet to be identified.