InnovEyes视力地图平台用于比较光线追踪引导的准分子激光原位角膜磨镶术(LASIK)和地形图引导的LASIK的研究。

Study of the InnovEyes Sightmap Platform in Comparing Ray-Tracing-Guided LASIK and Topography-Guided LASIK.

作者信息

Cao Xinfang, Zhang Jun, Shao Jie, Zhang Yonggang, Zheng Li

出版信息

J Refract Surg. 2024 Dec;40(12):e994-e1002. doi: 10.3928/1081597X-20241030-03. Epub 2024 Dec 1.

Abstract

PURPOSE

To evaluate the performance of the InnovEyes Sightmap platform (Alcon Laboratories, Inc) in refractive surgery by comparing the visual acuity and higher order aberrations (HOAs) between ray-tracing-guided laser in situ keratomileusis (LASIK) and topography-guided LASIK.

METHODS

This prospective study enrolled participants who underwent either ray-tracing-guided LASIK or topography-guided LASIK. Comprehensive ophthalmic evaluations were performed preoperatively, as well as at 1 day, 2 weeks, 1 month, and 3 months postoperatively. Patients in the ray-tracing-guided LASIK group underwent wavefront, tomography, and biometry assessment using the InnovEyes Sightmap diagnostic device. Assessments included visual acuity, manifest refraction, and whole-eye HOAs.

RESULTS

A total of 42 eyes treated with ray-tracing-guided LASIK and 42 eyes treated with topography-guided LASIK were analyzed. Both strategies demonstrated comparable good refraction accuracy and refractive stability ( > .05). The ray-tracing-guided LASIK group exhibited significantly better postoperative uncorrected distance visual acuity (UDVA) compared to the topography-guided LASIK group (-0.12 ± 0.05 vs -0.07 ± 0.04 logarithm of the minimum angle of resolution, respectively; < .05), with 48% of eyes achieving a UDVA of 20/12.5 or better. Ray-tracing-guided LASIK induced a small but statistically significant increase in HOAs and vertical coma aberration, along with a significant reduction in spherical aberration ( < .05). In contrast, topography-guided LASIK resulted in a significant increase in vertical coma ( < .05) without significant changes in overall HOAs or spherical aberration ( > .05). At 3 months postoperatively, spherical aberration was significantly different between the two groups (-0.021 ± 0.031 vs 0.054 ± 0.122 µm, respectively; < .05).

CONCLUSIONS

The InnovEyes Sightmap platform's ray-tracing-guided LASIK demonstrated potential advantages in visual acuity outcomes compared to topography-guided LASIK. The observed negative shift in spherical aberration, characterized by a lower absolute value, may have contributed to the enhanced visual acuity results. .

摘要

目的

通过比较光线追踪引导的准分子原位角膜磨镶术(LASIK)和地形图引导的LASIK之间的视力和高阶像差(HOA),评估InnovEyes Sightmap平台(爱尔康实验室公司)在屈光手术中的性能。

方法

这项前瞻性研究纳入了接受光线追踪引导的LASIK或地形图引导的LASIK的参与者。术前以及术后1天、2周、1个月和3个月进行了全面的眼科评估。光线追踪引导的LASIK组患者使用InnovEyes Sightmap诊断设备进行了波前、断层扫描和生物测量评估。评估包括视力、显验光和全眼HOA。

结果

共分析了42只接受光线追踪引导的LASIK治疗的眼睛和42只接受地形图引导的LASIK治疗的眼睛。两种策略均显示出相当好的屈光准确性和屈光稳定性(P>0.05)。与地形图引导的LASIK组相比,光线追踪引导的LASIK组术后裸眼远视力(UDVA)明显更好(分别为-0.12±0.05与-0.07±0.04最小分辨角对数;P<0.05),48%的眼睛达到了20/12.5或更好的UDVA。光线追踪引导的LASIK导致HOA和垂直彗差有小但具有统计学意义的增加,同时球差显著降低(P<0.05)。相比之下,地形图引导的LASIK导致垂直彗差显著增加(P<0.05),而总体HOA或球差无显著变化(P>0.05)。术后3个月,两组之间的球差有显著差异(分别为-0.021±0.031与0.054±0.122µm;P<0.05)。

结论

与地形图引导的LASIK相比,InnovEyes Sightmap平台的光线追踪引导的LASIK在视力结果方面显示出潜在优势。观察到的球差负向偏移,其特征是绝对值较低,可能有助于提高视力结果。