超声造影及微血流成像对乳腺癌患者血流信号检测的诊断价值初步研究

A preliminary study on the diagnostic value of contrast-enhanced ultrasound and micro-flow imaging for detecting blood flow signals in breast cancer patients.

作者信息

Dong Junjie, Chen Qiang, Wang Huiyang, He Hailing, Luo Ting, Jiang Tianan

机构信息

Department of Ultrasound Medicine, The First Affiliated Hospital, Zhejiang University School of Medicine, Hangzhou, China.

Department of Ultrasound Medicine, Taizhou Hospital, Zhejiang University, Taizhou, China.

出版信息

Gland Surg. 2024 Nov 30;13(11):2098-2106. doi: 10.21037/gs-24-264. Epub 2024 Nov 26.

Abstract

BACKGROUND

The distribution and morphology of mass microvessels could affect the diagnostic accuracy of breast cancer (BC). The aim of our study was to compare the value of contrast-enhanced ultrasound and micro-flow imaging (CEUS-MFI), contrast-enhanced ultrasound (CEUS), and color Doppler flow imaging (CDFI) in the assessment of mass microvasculature.

METHODS

A total of 106 patients with 106 breast masses categorized as Breast Imaging Reporting and Data System (BI-RADS) category 4 were enrolled in our prospective study. CEUS-MFI, CEUS and conventional CDFI were used to estimate the microvascular morphology and distribution types of breast lesions, respectively. Pathological results were considered the gold standard.

RESULTS

CEUS-MFI technique applied in microvascular morphology and distribution types resulted in a higher resolution in breast lesions than the CEUS and CDFI techniques. The sensitivity, specificity, and accuracy of CEUS-MFI were 94.4%, 85.3%, and 91.5%, respectively. The sensitivity, specificity, and accuracy of CEUS were 88.9%, 82.4%, and 86.8%, respectively. The sensitivity, specificity, and accuracy of the CDFI were 61.1%, 76.5% and 66.0%, respectively. There were significant differences in accuracy between the CEUS-MFI and CEUS (P=0.01). There were significant differences in accuracy between the CEUS-MFI and CDFI (P<0.001).

CONCLUSIONS

This study supports CEUS-MFI is a new and promising imaging method for visualizing microvasculature in breast masses. CEUS-MFI improves the diagnostic capacity for BC.

摘要

背景

乳腺肿块微血管的分布和形态可能会影响乳腺癌(BC)的诊断准确性。本研究旨在比较超声造影与微血管成像(CEUS-MFI)、超声造影(CEUS)和彩色多普勒血流成像(CDFI)在评估乳腺肿块微血管方面的价值。

方法

本前瞻性研究共纳入106例患有106个乳腺肿块的患者,这些肿块被分类为乳腺影像报告和数据系统(BI-RADS)4类。分别使用CEUS-MFI、CEUS和传统CDFI来评估乳腺病变的微血管形态和分布类型。病理结果被视为金标准。

结果

在乳腺病变的微血管形态和分布类型评估中,CEUS-MFI技术比CEUS和CDFI技术具有更高的分辨率。CEUS-MFI的敏感性、特异性和准确性分别为94.4%、85.3%和91.5%。CEUS的敏感性、特异性和准确性分别为88.9%、82.4%和86.8%。CDFI的敏感性、特异性和准确性分别为61.1%、76.5%和66.0%。CEUS-MFI与CEUS之间的准确性存在显著差异(P=0.01)。CEUS-MFI与CDFI之间的准确性存在显著差异(P<0.001)。

结论

本研究支持CEUS-MFI是一种用于可视化乳腺肿块微血管的新型且有前景的成像方法。CEUS-MFI提高了乳腺癌的诊断能力。