Tian Suqing, Sun Haitao, Zhao Tiandi, Wang Wei
Department Of Radiation Oncology, Peking University Third Hospital, Beijing 100191, China.
Beijing Da Xue Xue Bao Yi Xue Ban. 2024 Dec 18;56(6):1097-1100. doi: 10.19723/j.issn.1671-167X.2024.06.024.
To evaluate the intra-fraction and inter-fraction positional deviations in head and neck tumor patients undergoing intensity-modulated radiation therapy (IMRT) guided by cone-beam CT (CBCT), as well as the correction capability and stability of the HexaPOD evo RT 6D couch in addressing these deviations.
From May 2019 to April 2022, 59 consecutive patients with head and neck tumors were enrolled at the Department of Radiation Oncology, Peking University Third Hospital.Using the Elekta AXESSE image-guided stereotactic treatment system, a pre-treatment CBCT scan was performed, followed by bone window mode registration with the planning reference images.Deviations were corrected automatically or manually on the 6D couch, followed by a second CBCT scan for confirmation of the deviation correction.Positional errors in translation (, , directions) and rotation (, , directions) were recorded before and after correction, and intra-fraction and inter-fraction positional errors were analyzed.
Positional error data before and after correction of the 6D couch were successfully obtained and corrected online in 506 CBCT scans.The maximum positional errors before and after correction were 0.90 cm to 0.04 cm ( direction), 1.74 cm to 0.09 cm ( direction), 1.80 cm to 0.09 cm ( direction), and 2.90° to 0.14°( direction), 3.00° to 0.15°( direction), 3.00° to 0.15°( direction), respectively.The mean absolute values of translational (, , directions) and rotational (, , directions) errors significantly decreased after online correction, from 0.18 cm, 0.22 cm, 0.25 cm, and 0.82°, 1.11°, 0.73° to 0.01 cm, 0.01 cm, 0.01 cm, and 0.04°, 0.06°, 0.04°(all values < 0.001).After correction, the frequencies of translational errors less than 0.10 cm in the , , directions were 99.60%, 98.62%, and 95.45%, respectively, and the frequencies of rotational errors less than 0.2° were all above 99.80%.
Online correction combined with CBCT and the 6D couch significantly reduces both translational and rotational positional errors in patients undergoing head and neck radiation therapy, greatly enhancing the precision of treatment.
评估在锥形束CT(CBCT)引导下接受调强放射治疗(IMRT)的头颈部肿瘤患者的分次内和分次间位置偏差,以及HexaPOD evo RT 6D治疗床在解决这些偏差方面的校正能力和稳定性。
2019年5月至2022年4月,北京大学第三医院放疗科连续纳入59名头颈部肿瘤患者。使用医科达AXESSE图像引导立体定向治疗系统,治疗前进行CBCT扫描,然后与计划参考图像进行骨窗模式配准。在6D治疗床上自动或手动校正偏差,随后进行第二次CBCT扫描以确认偏差校正。记录校正前后在平移(x、y、z方向)和旋转(Pitch、Roll、Yaw方向)上的位置误差,并分析分次内和分次间位置误差。
成功获取了6D治疗床校正前后的位置误差数据,并在506次CBCT扫描中进行了在线校正。校正前后的最大位置误差分别为:x方向0.90 cm至0.04 cm,y方向1.74 cm至0.09 cm,z方向1.80 cm至0.09 cm,Pitch方向2.90°至0.14°,Roll方向3.00°至0.15°,Yaw方向3.00°至0.15°。在线校正后,平移(x、y、z方向)和旋转(Pitch、Roll、Yaw方向)误差的平均绝对值显著降低,从0.18 cm、0.22 cm、0.25 cm和0.82°、1.11°、0.73°降至0.01 cm、0.01 cm、0.01 cm和0.04°、0.06°、0.04°(所有P值<0.001)。校正后,x、y、z方向平移误差小于0.10 cm的频率分别为99.60%、98.62%和95.45%,旋转误差小于0.2°的频率均高于99.80%。
CBCT与6D治疗床相结合的在线校正显著降低了头颈部放射治疗患者的平移和旋转位置误差,大大提高了治疗精度。