宫腔镜子宫肌瘤切除术中Iglesias宫腔镜切除器喙部丢失的病例报告。

A case report on the loss of an Iglesias resectoscope beak during hysteroscopic myomectomy.

作者信息

Indraccolo Ugo

机构信息

Maternal and Child Department, Complex Operative Unit of Obstetrics and Gynecology, Hospital of Gubbio and Gualdo Tadino, ASL 1 Umbria, Perugia, PG, Italy.

出版信息

J Minim Access Surg. 2025 Jul 1;21(3):303-305. doi: 10.4103/jmas.jmas_376_23. Epub 2024 Dec 24.

Abstract

This case report describes the dynamics of the detachment of the ceramic beak of an Iglesias resectoscope that was lost in the uterine cavity during 'cold loop' hysteroscopic myomectomy. Our aim is to increase awareness of this issue and caution our colleagues against using Iglesias resectoscopes for hysteroscopies. The ceramic beak of an Iglesias resectoscope can detach itself if hard tissue is encountered during operative hysteroscopy.

摘要

本病例报告描述了在“冷环”宫腔镜子宫肌瘤切除术中,伊格莱西亚斯子宫切除器的陶瓷喙在子宫腔内脱落的动态过程。我们的目的是提高对这个问题的认识,并告诫我们的同事不要使用伊格莱西亚斯子宫切除器进行宫腔镜检查。如果在手术宫腔镜检查中遇到硬组织,伊格莱西亚斯子宫切除器的陶瓷喙可能会自行脱落。