Clinical Validation of M-TONX: A Novel Combo Rebound Tonometer and Pachymeter.

作者信息

Nilforushan Naveed, Yousefi Mozafar, Babic Ankica, Gharehbaghi Arash

机构信息

Eye Research Centre, Iran University of Medical Sciences, Tehran, Iran.

Moj Teb Parsian Incorporation, Tehran, Iran.

出版信息

Transl Vis Sci Technol. 2024 Dec 2;13(12):34. doi: 10.1167/tvst.13.12.34.

Abstract

PURPOSE

This study aims to perform a clinical investigation of an innovative rebound technology-based device, the M-TONX, to simultaneously measure intraocular pressure (IOP) and central corneal thickness (CCT).

METHODS

The IOP and CCT of the patients were first measured by the M-TONX. Then, the measurements were repeated by the Goldman applanation (GAT) and the Pentacam corneal topographer, as the standard devices. For the statistical analysis, the patients were stratified based on their IOPs to group 1 (IOP < = 16 millimeters of mercury [mm Hg]), group 2 (16 mm Hg < IOP < 23 mm Hg), and group 3 (IOP > = 23 mm Hg). The stratification was also performed for the CCTs as: class 1 (CCT < = 475 µm), class 2 (475 < CCT < 574 µm), and class 3 (CCT > = 575 µm).

RESULTS

Of the 374 eyes (225 subjects and 43% women), 262, 66, and 46 eyes belonged to group 1, group 2, and group 3, respectively. Very high IOP (>35 mm Hg) was observed in 12 eyes. The overall confidence interval of the deviation (confidence level = 95%) from the standard devices was estimated to be 0.7 to 1.2 mm Hg for IOP, and -12.7 to -5.4 µm for CCT. The stratified analysis showed substantial agreement with the standard devices with the intraclass correlation > 0.65 and the Pearson Correlation > 0.8 calculated for all the groups and classes.

CONCLUSIONS

The M-TONX exhibited a reliable performance concerning the standards for measuring IOP and CCT. Its accuracy remains stable for a broad range of IOP and CCT. The M-TONX successfully incorporates two separate functionalities into a single compact user-friendly device.

TRANSLATIONAL RELEVANCE

This study uncovers the conformity of the technology with the standards, linking fundamental research to clinical care.

摘要

目的

本研究旨在对一种基于创新回弹技术的设备M-TONX进行临床研究,以同时测量眼压(IOP)和中央角膜厚度(CCT)。

方法

首先使用M-TONX测量患者的IOP和CCT。然后,使用戈德曼压平眼压计(GAT)和Pentacam角膜地形图仪作为标准设备重复测量。为了进行统计分析,根据患者的IOP将其分为1组(IOP≤16毫米汞柱[mmHg])、2组(16 mmHg<IOP<23 mmHg)和3组(IOP≥23 mmHg)。对CCT也进行了分层:1级(CCT≤475 µm)、2级(475<CCT<574 µm)和3级(CCT≥575 µm)。

结果

在374只眼中(225名受试者,43%为女性),分别有262只、66只和46只眼属于1组、2组和3组。在12只眼中观察到眼压非常高(>35 mmHg)。与标准设备相比,IOP偏差的总体置信区间(置信水平=95%)估计为0.7至1.2 mmHg,CCT为-12.7至-5.4 µm。分层分析显示,与标准设备具有高度一致性,所有组和级别计算的组内相关性>0.65,皮尔逊相关性>0.8。

结论

M-TONX在测量IOP和CCT的标准方面表现出可靠的性能。其准确性在广泛的IOP和CCT范围内保持稳定。M-TONX成功地将两种独立的功能整合到一个紧凑、用户友好的单一设备中。

转化相关性

本研究揭示了该技术与标准的一致性,将基础研究与临床护理联系起来。