Donald and Barbara Zucker School of Medicine at Hofstra Northwell, Hempstead.
Northwell Health North Shore-LIJ Department of Ophthalmology, Great Neck, NY.
J Glaucoma. 2018 Dec;27(12):1119-1124. doi: 10.1097/IJG.0000000000001059.
To compare intraocular pressure (IOP) measurements by Goldmann applanation tonometer (GAT), disposable Goldmann applanation prism, ICare, and Tonopen.
A total of 74 patients with varying glaucoma status were examined in our outpatient clinic and IOP was measured with 3 tonometers. The disposable Tonojet prism (dGAT), Tonopen XL, and ICare ic100 were compared with the GAT.
There were good intraclass correlation coefficients between IOP measurements by GAT and dGAT (0.95), Tonopen (0.83), and ICare (0.77), all P<0.001. The IOP mean differences between dGAT and GAT were mean 0.80 mm Hg; 95% limits of agreement: -3.35 to 4.96 mm Hg. For Tonopen and GAT: mean, -1.67 mm Hg; limits of agreement, -8.55 to 5.21 mm Hg. For ICare and GAT: mean, 0.44 mm Hg; limits of agreement, -8.18 to 9.06 mm Hg.
The most reliable modality, with good correlation with the Goldmann tonometer values, was the GAT with dGAT, followed in descending order by the Tonopen XL and ICare. There was good interdevice agreement and consistency between all devices. On subgroup analysis, all 3 modalities were found to be less reliable at extreme IOP values (<10 and >24 mm Hg). These disposable modalities should be avoided in extreme IOP ranges outside the normal range.
比较 Goldmann 压平眼压计(GAT)、一次性 Goldmann 压平棱镜、ICare 和 Tonopen 的眼压测量值。
我们在门诊检查了 74 名不同青光眼患者,并使用 3 种眼压计测量眼压。一次性 Tonojet 棱镜(dGAT)、Tonopen XL 和 ICare ic100 与 GAT 进行比较。
GAT 与 dGAT(0.95)、Tonopen(0.83)和 ICare(0.77)之间的眼压测量值具有良好的组内相关系数,均 P<0.001。dGAT 与 GAT 的眼压平均差异为 0.80mmHg;95%一致性界限:-3.35 至 4.96mmHg。对于 Tonopen 和 GAT:平均差异为-1.67mmHg;一致性界限:-8.55 至 5.21mmHg。对于 ICare 和 GAT:平均差异为 0.44mmHg;一致性界限:-8.18 至 9.06mmHg。
最可靠的模式,与 Goldmann 眼压计值具有良好的相关性,是带有一次性 dGAT 的 GAT,其次是 Tonopen XL 和 ICare。所有设备之间的设备间一致性和一致性良好。在亚组分析中,所有 3 种模式在眼压值极值(<10 和>24mmHg)时均不太可靠。在正常范围之外的极端眼压范围内,应避免使用这些一次性设备。