Starritt H C, Perkins M A, Duck F A, Humphrey V F
J Acoust Soc Am. 1985 Jan;77(1):302-6. doi: 10.1121/1.392272.
Finite-amplitude distortion of ultrasonic waves from medical equipment has been observed to occur following transmission through calf muscle in human volunteers. Measurements were made using both dynamic pulse-echo imaging equipment and physiotherapy equipment. In both cases irradiation was carried out under operating conditions commonly used clinically. Pressure waveforms were measured at the skin surface using a broadband polyvinylidene difluoride membrane hydrophone. Using a pulsed, weakly focused 2.5-MHz beam with input peak pressure of 0.8 MPa and a pressure gain of 5.3 at the focus, the mean second harmonic peak magnitude (16 measurements) was 17 dB below the fundamental peak. A 1.1-MHz continuous wave therapy set with input peak pressure of 0.5 MPa showed mean second harmonic magnitude 23 dB below the fundamental.
人们观察到,医疗设备发出的超声波在通过人类志愿者的小腿肌肉传输后会出现有限振幅失真。使用动态脉冲回波成像设备和物理治疗设备进行了测量。在这两种情况下,照射均在临床上常用的操作条件下进行。使用宽带聚偏二氟乙烯薄膜水听器在皮肤表面测量压力波形。使用脉冲、弱聚焦的2.5兆赫波束,输入峰值压力为0.8兆帕,焦点处压力增益为5.3,平均二次谐波峰值幅度(16次测量)比基波峰值低17分贝。一台输入峰值压力为0.5兆帕的1.1兆赫连续波治疗仪显示,平均二次谐波幅度比基波低23分贝。