一种使用多参数优化技术为乳腺摄影指定最小剂量系统的方法。

An approach to specifying a minimum dose system for mammography using multiparameter optimization techniques.

作者信息

Muntz E P, Jafroudi H, Jennings R, Bernstein H

出版信息

Med Phys. 1985 Jan-Feb;12(1):5-12. doi: 10.1118/1.595737.

Abstract

Analytical expressions have been written for image quality in mammography. Multiparameter optimizations have been conducted to find mammography systems providing the lowest patient dose for a given image quality. The optimizations are subject to constraints imposed by technology, such as power limits on the tube focal spot, absorption efficiency related to detector resolution, and others. The optimizations permit system geometry, kVp, filtration, detector resolution, focal spot size, and grid characteristics to vary simultaneously and self-consistently subject to the constraints. A system configuration approaching a factor of 3 dose reduction has been found without assuming radical technological advances. The system satisfies image quality constraints for both large and small targets and would be possible to implement clinically. The sensitivity of the results to the assumptions made in the modeling has been investigated.

摘要

已针对乳腺摄影中的图像质量写出了分析表达式。已进行多参数优化,以找到在给定图像质量下为患者提供最低剂量的乳腺摄影系统。这些优化受到技术所施加的限制,例如管焦点的功率限制、与探测器分辨率相关的吸收效率等。这些优化允许系统几何结构、千伏峰值(kVp)、过滤、探测器分辨率、焦点尺寸和滤线栅特性在这些限制条件下同时且自洽地变化。在不假设激进技术进步的情况下,已找到一种剂量降低近3倍的系统配置。该系统满足大小目标的图像质量约束,并且有可能在临床上实施。已研究了结果对建模中所做假设的敏感性。