Validation of glucose and lactate in cerebrospinal fluid (CSF) on a Radiometer blood gas analyzer ABL90 Flex plus.

作者信息

Thakur Vinita, Akerele Olatunji Anthony, Randell Edward

机构信息

Pathology and Laboratory Medicine Program, Health Sciences Centre, St. John's, Newfoundland and Labrador, Canada; Memorial University of Newfoundland, Health Sciences Centre, St. John's, Newfoundland and Labrador, Canada.

Pathology and Laboratory Medicine Program, Health Sciences Centre, St. John's, Newfoundland and Labrador, Canada.

出版信息

Clin Biochem. 2025 Mar;136:110876. doi: 10.1016/j.clinbiochem.2025.110876. Epub 2025 Jan 5.

Abstract

PURPOSE

Rapid determination of cerebrospinal fluid (CSF) glucose and lactate is required by emergency rooms and intensive care units. Long turnaround time (TAT) on test results negatively impacts timely diagnosis and treatment of neurological infections like meningitis.

METHODS

The CSF glucose and lactate assays were evaluated on a blood gas analyzer, Radiometer ABL90 Flex Plus. Linearity, limit of quantitation (LOQ), and precision were determined using fresh and spiked patient CSF samples. Fifty-four fresh and 49 frozen CSF samples were used to compare the method againstAbbottArchitectC16000. An inter-laboratory comparison was done across eight hospital sites having ABL90 Flex Plus. The stability of both tests was tested for 48 h at ambient and refrigerated temperatures. Results were compared between centrifuged and uncentrifuged fresh CSF samples to determine if particulate in uncentrifuged samples impacted analysis.

RESULTS

Glucose and lactate assays were linear over a broad analytical range of 1-45 mmol/L and 0-37 mmol/L, respectively, and demonstrated a good correlation with the routine chemistry laboratory method. LOQ was determined as 0.4 mmol/L for CSF glucose with a coefficient of variation (CV) of 14.7 % and 0.2 mmol/L with 0 % CV for CSF lactate respectively. Repeatability and reproducibility show small imprecision for both these assays. Glucose and lactate were stable for over 48 h at room or refrigeration temperatures. Sample particulates had no impact on the measurement. The inter-laboratory comparison was within total allowable error for glucose and lactate.

CONCLUSIONS

Acceptable performance characteristics, small sample volume, and rapid TAT make ABL90 Flex Plus an acceptable alternative analyzer for CSF glucose and lactate.