Suppr超能文献

Three-way comparison of retinoscopy, subjective, and Dioptron Nova refractive findings.

作者信息

Grosvenor T, Perrigin D M, Perrigin J

出版信息

Am J Optom Physiol Opt. 1985 Jan;62(1):63-5. doi: 10.1097/00006324-198501000-00009.

Abstract

On the basis of routine clinical findings taken on a group of 100 patients serving as subjects in the Houston Myopia Control study, a three-way comparison was made between retinoscopy, subjective, and Dioptron Nova refractive findings. Retinoscopy and subjective findings compared very favorably, mean differences being -0.01 D for spherical equivalent power and -0.05 D for cylinder power. Dioptron and retinoscopy findings differed by -0.30 D for spherical equivalent power and -0.13 D for cylinder power, whereas Dioptron and subjective findings differed by -0.33 D for spherical equivalent power and -0.18 D for cylinder power.