Objective Approach for Titration of Continuous Positive Airway Pressure in Patients with Obstructive Sleep Apnea Using Daytime Continuous Positive Airway Pressure Titration Based on Respiratory Movement Regularity.

作者信息

Burioka Naoto, Hirayama Yuki, Endo Masahiro, Oguri Masayoshi, Takata Miyako, Ikeuchi Tomoyuki, Yamasaki Akira

机构信息

Department of Respiratory Medicine, Yonago Medical Center, National Hospital Organization, Yonago 683-0006, Japan.

Department of Pathobiological Science and Technology, Graduate School of Medical Sciences, School of Health Science, Faculty of Medicine, Tottori University, Yonago 683-8503, Japan.

出版信息

J Clin Med. 2024 Dec 13;13(24):7603. doi: 10.3390/jcm13247603.

Abstract

: Continuous positive airway pressure (CPAP) is used to treat patients with obstructive sleep apnea (OSA) and has proven clinical efficacy for this condition. However, the objective method to determine the appropriate CPAP level for treatment is still unclear. Patients with OSA typically exhibit irregular respiratory efforts due to obstruction or narrowing of the upper airway during sleep. Treatment with an adequate CPAP level alleviates airway obstruction or narrowing, leading to more regular respiratory patterns. We aimed to develop an objective CPAP titration method using the respiratory regularity index (RRI) derived from respiratory movements. : We conducted daytime CPAP titration in 10 patients with OSA. Respiratory movements were recorded by inductance plethysmography in three conditions: wake, apnea, and during CPAP therapy. The RRI of respiratory movements was calculated in each condition, and the CPAP level with the lowest RRI was identified as the appropriate air pressure level in each patient. CPAP therapy at the appropriate level determined by the study method was conducted for 2 months, after which efficacy was assessed by night polysomnography in the hospital. : The fixed air pressure level for the CPAP device was determined as 7.8 ± 0.7 cmHO based on our daytime CPAP titration method using the RRI. The apnea-hypopnea index improved significantly from before CPAP therapy (43.1 ± 15.3 h) to during CPAP therapy (3.0 ± 2.1 h) ( < 0.006). The Epworth Sleepiness Scale score decreased significantly during CPAP therapy. The lowest percutaneous arterial oxygen saturation values in each sleep stage also improved significantly with the CPAP determined by our daytime CPAP titration method. : The study findings showed that our daytime CPAP titration method based on the RRI is an effective method for determining individualized appropriate CPAP levels in patients with OSA. The RRI-based daytime CPAP titration method offers a simplified approach to determining the adequate CPAP level for patients with OSA.

摘要

持续气道正压通气(CPAP)用于治疗阻塞性睡眠呼吸暂停(OSA)患者,并且已证实对该病症具有临床疗效。然而,确定合适的CPAP治疗水平的客观方法仍不明确。OSA患者在睡眠期间通常由于上气道阻塞或狭窄而表现出不规则的呼吸努力。使用适当水平的CPAP进行治疗可缓解气道阻塞或狭窄,从而产生更规律的呼吸模式。我们旨在开发一种使用源自呼吸运动的呼吸规律指数(RRI)的客观CPAP滴定方法。

我们对10例OSA患者进行了日间CPAP滴定。通过感应体积描记法在三种状态下记录呼吸运动:清醒、呼吸暂停和CPAP治疗期间。在每种状态下计算呼吸运动的RRI,并将具有最低RRI的CPAP水平确定为每位患者的合适气压水平。采用本研究方法确定的合适水平进行CPAP治疗2个月,之后在医院通过夜间多导睡眠图评估疗效。

基于我们使用RRI的日间CPAP滴定方法,CPAP设备的固定气压水平确定为7.8±0.7 cmH₂O。呼吸暂停低通气指数从CPAP治疗前(43.1±15.3次/小时)到CPAP治疗期间(3.0±2.1次/小时)显著改善(P<0.006)。在CPAP治疗期间,爱泼沃斯嗜睡量表评分显著降低。采用我们的日间CPAP滴定方法确定的CPAP治疗后,每个睡眠阶段的最低经皮动脉血氧饱和度值也显著改善。

研究结果表明,我们基于RRI的日间CPAP滴定方法是确定OSA患者个体化合适CPAP水平的有效方法。基于RRI的日间CPAP滴定方法为确定OSA患者的合适CPAP水平提供了一种简化方法。