Solé Morillo Ángel, Cause Joan Lambert, De Pauw Kevin, Silva Bruno da, Stiens Johan
Department of Electronics and Informatics (ETRO), Vrije Universiteit Brussel (VUB), 1050 Brussels, Belgium.
Department of Biomedical Engineering, Universidad de Oriente, Santiago de Cuba 90500, Cuba.
Sensors (Basel). 2024 Dec 27;25(1):99. doi: 10.3390/s25010099.
Photoplethysmography is a widely used optical technique to extract physiological information non-invasively. Despite its large use and adoption, multiple factors influence the signal shape and quality, including the instrumentation used. This work analyzes the variability of the DC component of the PPG signal at three source-detector distances (6 mm, 9 mm, and 12 mm) using green, red, and infrared light and four photodiodes per distance. The coefficient of variation (CV) is proposed as a new signal quality index (SQI) to evaluate signal variabilities. This study first characterizes the PPG system, which is then used to acquire PPG signals in the chest of 14 healthy participants. Results show a great DC variability at 6 mm, homogenizing at 9 and 12 mm. This suggests that PPG systems are also sensitive to the near- and far-field effects commonly reported and studied in optics, which can impact the accuracy of physiological parameters dependent on the DC component, such as oxygen saturation (SpO).
光电容积脉搏波描记法是一种广泛应用的光学技术,用于无创提取生理信息。尽管其被大量使用和采用,但多种因素会影响信号形状和质量,包括所使用的仪器。这项工作分析了在三个源 - 探测器距离(6毫米、9毫米和12毫米)下,使用绿色、红色和红外光以及每个距离四个光电二极管时,PPG信号直流分量的变化性。提出变异系数(CV)作为一种新的信号质量指标(SQI)来评估信号变化性。本研究首先对PPG系统进行了表征,然后用于采集14名健康参与者胸部的PPG信号。结果表明,在6毫米处直流变化很大,在9毫米和12毫米处趋于均匀。这表明PPG系统对光学中常见的近场和远场效应也很敏感,这可能会影响依赖于直流分量的生理参数的准确性,如血氧饱和度(SpO)。