Comparison of a modified manual hexadimethrine bromide (Polybrene) and a low-ionic-strength solution antibody detection technique.

作者信息

Ferrer Z, Wright J, Moore B P, Freedman J

出版信息

Transfusion. 1985 Mar-Apr;25(2):145-8. doi: 10.1046/j.1537-2995.1985.25285169208.x.

Abstract

Manual hexadimethrine bromide (Polybrene) tests (Polybrene in low-ionic medium) were used in parallel with manual low-ionic-strength solution (LISS) procedures for the routine testing of patient samples referred to a general hospital blood bank. Of 5646 consecutive sera tested, 5167 (91.5%) did not react with either technique; 320 sera (5.7%) reacted in both methods. The Polybrene technique detected 63 antibodies which did not react in the LISS methods. One hundred sera did not react in the Polybrene test, but did react in the LISS methods. Sera showing discrepant results between the two methods were further tested in a reference laboratory. Polybrene tests appeared to be better in avoiding reactions due to clinically nonsignificant antibodies. The LISS methods, however, appeared to be more sensitive in detecting antibodies of potential clinical significance.

摘要

相似文献

4
Limitations of Polybrene to detect ABO incompatibility.
Vox Sang. 1986;51(4):318-20. doi: 10.1111/j.1423-0410.1986.tb01976.x.
7
The manual hexadimethrine bromide (Polybrene) test. Effects of serum proteins and practical applications.
Transfusion. 1987 Jul-Aug;27(4):295-301. doi: 10.1046/j.1537-2995.1987.27487264734.x.