An ion-exchange chromatographic procedure is described which facilitates the determination of beta-chain aminoterminal modified glycated hemoglobin. The procedure includes an erythrocyte lysis reagent which eliminates the labile aldimine component (pre-A1c) and a two-stage elution step which separates HbA, 1a + b from HbA1c. This procedure also includes calibrator material which aids in correcting for temperature fluctuations during the analysis. Within-run CV's for samples with HbA1c levels between 4.0% and 13.7% were 1.4 to 3.2%. The between-run CV for an HbA1c control was 5.5%. A comparison of the present test to an ion-exchange HPLC method yielded the equation: HPLC = 0.96 (present method) -0.2% (n = 101 and r = 0.984). Two separate reference range studies yielded comparable results (n = 220/65, mean = 4.77/4.78%, S.D. = 0.68/0.55). Studies with pooled erythrocytes and various lipemic plasmas did not reveal any assay interferences. Various abnormal hemoglobins were studied for their effect on the assay.

摘要

本文描述了一种离子交换色谱法,该方法有助于测定β链氨基末端修饰的糖化血红蛋白。该方法包括一种红细胞裂解试剂,可消除不稳定的醛亚胺成分(前A1c),以及一个两步洗脱步骤,可将HbA、1a + b与HbA1c分离。该方法还包括校准材料,有助于校正分析过程中的温度波动。HbA1c水平在4.0%至13.7%之间的样品批内变异系数为1.4%至3.2%。HbA1c对照品的批间变异系数为5.5%。将本试验与离子交换HPLC法进行比较,得出方程:HPLC = 0.96(本方法)-0.2%(n = 101,r = 0.984)。两项独立的参考范围研究得出了可比的结果(n = 220/65,平均值 = 4.77/4.78%,标准差 = 0.68/0.55)。对混合红细胞和各种脂血血浆的研究未发现任何分析干扰。研究了各种异常血红蛋白对该分析的影响。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验