Abstract

OBJECTIVE

The aim of this study was to perform a comparative pre-clinical evaluation of a new fenestrated endovascular aneurysm repair dedicated stent using perfused 3D printed patient anatomies.

METHODS

The test setup included a pulsatile pump set to reproduce human haemodynamics, four 3D printed pararenal aneurysms connected to a bench test, and four corresponding fenestrated grafts (Zenith Fenestrated; Cook, Bloomington, IN, USA). Bridging stents were sized based on analysis of patients computed tomography scans and included either four standard covered stents (BeGraft [BG]; Bentley, Hechingen, Germany) or four similar covered stent grafts mounted on a modified balloon to allow stent implantation and flaring in one step (BeFlared [BF]; Bentley). Each anatomy and corresponding endograft was used for four sets of four bridging stent implantations. Selective angiography following each stent implantation was used to assess efficacy (patency and absence of type Ic or IIIc endoleak), procedure time was recorded, and cone beam computed tomography at the end of each set of four stent implantations enabled comparison of geometric layout between stents. Physician assessment evaluated technical performance and failure modes.

RESULTS

Sixty four stents were successfully implanted. Selective angiograms depicted three type IIIc (two in BG, one in BF) and seven type Ic (four in BG, three in BF) endoleaks, which were all corrected intra-operatively. Compared with BG, the BF had a larger mean proximal diameter inside the endograft (10.5 mm vs. 9.2 mm; p < .001) and greater opening flaring angle (46.1° vs. 33.4°; p < .001). The eight procedures performed with the BF were quicker than BG (17.0 ± 4 minutes vs. 23.9 ± 4.7 minutes; p = .004). Physician assessment did not reveal any failure mode and concluded an excellent technical performance of both devices.

CONCLUSION

The BF demonstrated similar efficacy and performance compared with the BG for bridging target vessels in an innovative test setup using 3D printed patient anatomies.

摘要

目的

本研究旨在使用灌注式3D打印患者解剖模型对一种新型开窗型血管内动脉瘤修复专用支架进行临床前对比评估。

方法

测试装置包括一台设置为模拟人体血流动力学的搏动泵、四个连接到台架试验的3D打印肾旁动脉瘤,以及四个相应的开窗型移植物(Zenith开窗型;库克公司,美国印第安纳州布卢明顿)。根据患者计算机断层扫描分析确定桥接支架尺寸,包括四个标准覆膜支架(BeGraft [BG];本特利公司,德国黑兴根)或四个安装在改良球囊上的类似覆膜支架移植物,以便一步完成支架植入和扩张(BeFlared [BF];本特利)。每个解剖模型和相应的腔内移植物用于四组,每组进行四次桥接支架植入。每次支架植入后进行选择性血管造影以评估疗效(通畅性以及无Ic型或IIIc型内漏),记录手术时间,并且在每组四次支架植入结束时进行锥形束计算机断层扫描,以便比较支架之间的几何布局。医生评估技术性能和失败模式。

结果

成功植入64个支架。选择性血管造影显示三个IIIc型内漏(两个在BG组,一个在BF组)和七个Ic型内漏(四个在BG组,三个在BF组),所有内漏均在术中得到纠正。与BG相比,BF在腔内移植物内的近端平均直径更大(10.5毫米对9.2毫米;p < 0.001),开口扩张角度更大(46.1°对33.4°;p < 0.001)。使用BF进行的八次手术比BG更快(17.0 ± 4分钟对23.9 ± 4.7分钟;p = 0.004)。医生评估未发现任何失败模式,并得出两种装置技术性能均优异的结论。

结论

在使用3D打印患者解剖模型的创新测试装置中,BF在桥接目标血管方面与BG显示出相似的疗效和性能。