Abstract

SIGNIFICANCE

Laparoscopic surgery is generally unavailable in low- and middle-income countries (LMICs) due to the high cost of installation and lack of qualified personnel to maintain and repair equipment. We developed a low-cost, durable, reusable laparoscopic system, called the KeyScope laparoscope, for use in LMICs. To reliably build and service the KeyScope in LMICs, a portable testing chamber (PTC) is needed to assess image performance.

AIM

A PTC was developed to characterize KeyScope laparoscope performance in LMICs.

APPROACH

Images of standard resolution, color accuracy, distortion, and depth of field (DOF) targets were captured in both a standard optical bench setup (OBS) and the PTC. Measurements from the OBS and PTC were quantified and compared using standard software (ImageJ and Imatest). To further reduce cost, alternative paper imaging targets were identified and compared with standard glass targets. To improve usability, MATLAB applications (apps) were developed to automate image analysis and reduce cost.

RESULTS

The PTC achieved similar results compared to the OBS for the image quality metrics, distortion and DOF. Further, the PTC presented similar results to the OBS for resolution at 4 to 7 cm working distances and improved resolution at periphery working distances of 3 and 10 cm. Color accuracy values were also improved in the PTC compared with those measured in the OBS. The low-cost resolution, color accuracy, and distortion targets resulted in similar image quality results to the standard image quality target. MATLAB apps produced similar results to Imatest and ImageJ software and decreased the time to complete image quality test analysis.

CONCLUSION

The low-cost portable design of the PTC will facilitate the translation of the KeyScope by enabling accurate and fast characterization of laparoscopic imaging performance in LMICs.

摘要

意义

由于安装成本高昂且缺乏维护和修理设备的合格人员,腹腔镜手术在低收入和中等收入国家(LMICs)通常无法开展。我们开发了一种低成本、耐用、可重复使用的腹腔镜系统,称为KeyScope腹腔镜,用于LMICs。为了在LMICs可靠地制造和维护KeyScope,需要一个便携式测试腔(PTC)来评估图像性能。

目的

开发一种PTC,以表征KeyScope腹腔镜在LMICs中的性能。

方法

在标准光学平台设置(OBS)和PTC中均采集了标准分辨率、颜色准确性、失真和景深(DOF)目标的图像。使用标准软件(ImageJ和Imatest)对来自OBS和PTC的测量结果进行量化和比较。为了进一步降低成本,确定了替代纸质成像目标并与标准玻璃目标进行比较。为了提高可用性,开发了MATLAB应用程序(应用)以实现图像分析自动化并降低成本。

结果

对于图像质量指标、失真和DOF,PTC与OBS相比取得了相似的结果。此外,在4至7厘米工作距离处,PTC的分辨率与OBS相似,在3厘米和10厘米的周边工作距离处分辨率有所提高。与在OBS中测量的值相比,PTC中的颜色准确性值也有所提高。低成本的分辨率、颜色准确性和失真目标产生的图像质量结果与标准图像质量目标相似。MATLAB应用程序产生的结果与Imatest和ImageJ软件相似,并减少了完成图像质量测试分析的时间。

结论

PTC的低成本便携式设计将通过能够在LMICs中准确快速地表征腹腔镜成像性能,促进KeyScope的推广应用。