使用径向梯度和自旋回波(RADGRASE)脉冲序列对乳腺进行水T和脂肪分数的同步映射。

Concurrent water T and fat fraction mapping of the breast using the radial gradient and spin echo (RADGRASE) pulse sequence.

作者信息

Hagio Tomoe, Galons Jean-Philippe, Roe Denise, Marron Marylin T, Thomson Cynthia, Thompson Patricia, Stopeck Alison T, Bilgin Ali, Altbach Maria I, Chiang Jing-Tzyh Alan

机构信息

Department of Biomedical Engineering, The University of Arizona, Tucson, AZ 85721, USA.

Department of Medical Imaging, The University of Arizona, Tucson, AZ 85724, USA.

出版信息

Magn Reson Imaging. 2025 May;118:110355. doi: 10.1016/j.mri.2025.110355. Epub 2025 Feb 5.

Abstract

This work describes and evaluates an efficient radial gradient- and spin-echo (RADGRASE) pulse sequence and reconstruction algorithm for concurrent measurement of proton-density weighted fat fraction (FF) and water component T (T) within breast tissues. The ability to estimate T in breast tissues, where fat can be highly abundant, is demonstrated using oil/gel phantoms across a wide range of FF values (0.1-0.7). Successful T mapping of breast tissues is also demonstrated in vivo by comparison with fat suppressed T values. The sensitivity of RADGRASE to detect changes in the breast was assessed by tracking T in 3 healthy volunteers through their menstrual cycle, demonstrating T values in the late luteal phase to be 18-29 ms higher than in the follicular phase. The technique is also applied to a cohort of 68 patients taking tamoxifen for breast cancer risk reduction, where significant positive correlation between the FF parameter Frac50 and T (p = 0.035) was observed in premenopausal subjects (n = 20). Our findings demonstrate the ability and efficacy of RADGRASE for simultaneously mapping FF and T within breast tissues, and the potential utility of the technique in studying breast tissue changes in clinical applications.

摘要

这项工作描述并评估了一种用于同时测量乳腺组织中质子密度加权脂肪分数(FF)和水成分T1(T1)的高效径向梯度和自旋回波(RADGRASE)脉冲序列及重建算法。使用油/凝胶体模在广泛的FF值范围(0.1 - 0.7)内证明了在脂肪含量可能很高的乳腺组织中估计T1的能力。通过与脂肪抑制T1值比较,还在体内成功地对乳腺组织进行了T1映射。通过跟踪3名健康志愿者在月经周期中的T1值,评估了RADGRASE检测乳腺变化的敏感性,结果表明黄体晚期的T1值比卵泡期高18 - 29毫秒。该技术还应用于一组68名服用他莫昔芬以降低乳腺癌风险的患者,在绝经前受试者(n = 20)中观察到FF参数Frac50与T1之间存在显著正相关(p = 0.035)。我们的研究结果证明了RADGRASE在同时绘制乳腺组织中的FF和T1图谱方面的能力和功效,以及该技术在临床应用中研究乳腺组织变化的潜在实用性。

相似文献

2
A flexible fast spin echo triple-echo Dixon technique.一种灵活的快速自旋回波三回波狄克逊技术。
Magn Reson Med. 2017 Mar;77(3):1049-1057. doi: 10.1002/mrm.26186. Epub 2016 Mar 8.
7
Water-fat Dixon cardiac magnetic resonance fingerprinting.水脂分离 Dixon 心脏磁共振指纹成像
Magn Reson Med. 2020 Jun;83(6):2107-2123. doi: 10.1002/mrm.28070. Epub 2019 Nov 18.

本文引用的文献

1
Tamoxifen evolution.他莫昔芬的演变
Br J Cancer. 2023 Feb;128(3):421-425. doi: 10.1038/s41416-023-02158-5. Epub 2023 Feb 10.
5
Imaging Breast Density: Established and Emerging Modalities.乳腺密度成像:既定与新兴模式
Transl Oncol. 2015 Dec;8(6):435-45. doi: 10.1016/j.tranon.2015.10.002.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验