J Chromatogr. 1985 Apr 26;323(2):191-225. doi: 10.1016/s0021-9673(01)90384-x.
Unmodified silica columns together with non-aqueous ionic eluents give stable yet flexible systems for the analysis of basic drugs by high-performance liquid chromatography. Low-wavelength UV and fluorescence detection may be used, and fluorescence may be optimised by, for example, post-column pH change or derivatisation of some primary aliphatic amines with o-phthaldialdehyde. A novel feature is that electrochemical oxidation can be used for the detection of most analytes and this detection mode is thus discussed in detail. Retention and relative response data (UV, 254 nm and electrochemical, +1.2 V) have been generated for 462 compounds using a 125-mm Spherisorb S5W silica column and methanolic ammonium perchlorate (10 mM, pH 6.7) as eluent. This system can be used isocratically in qualitative analyses and also for quantitative work, when either the wavelength or the applied potential can be adjusted to optimise the response.
未改性硅胶柱与非水离子洗脱剂相结合,为高效液相色谱分析碱性药物提供了稳定且灵活的系统。可使用低波长紫外和荧光检测,荧光可通过例如柱后pH变化或用邻苯二甲醛对一些伯脂肪胺进行衍生化来优化。一个新颖的特点是,电化学氧化可用于大多数分析物的检测,因此将详细讨论这种检测模式。使用125mm Spherisorb S5W硅胶柱和甲醇高氯酸铵(10mM,pH 6.7)作为洗脱剂,已生成了462种化合物的保留和相对响应数据(紫外,254nm和电化学,+1.2V)。该系统可在定性分析中以等度方式使用,也可用于定量工作,此时可调整波长或施加的电位以优化响应。