Catheters for arterial pressure monitoring in pediatrics.

作者信息

Fiser D H, Graves S A, van der Aa J

出版信息

Crit Care Med. 1985 Jul;13(7):580-3. doi: 10.1097/00003246-198507000-00015.

Abstract

To evaluate the accuracy of an underdamped catheter transducer recorder used to measure arterial blood pressure, natural frequency and damping ratio were determined by transient testing with two types of 22-ga and 24-ga catheters. Damping coefficients were considerably higher and resonant frequencies lower with the 24-ga catheters. The difference between waveforms of known pressure and HR and the output waveforms produced by the catheter transducer systems was determined by using a BP systems calibrator. Catheter size did not affect the difference in recorded pressures.

摘要

为评估用于测量动脉血压的欠阻尼导管换能器记录仪的准确性,通过对两种型号的22号和24号导管进行瞬态测试来确定其固有频率和阻尼比。24号导管的阻尼系数明显更高,共振频率更低。使用血压系统校准仪确定已知压力和心率的波形与导管换能器系统产生的输出波形之间的差异。导管尺寸不影响记录压力的差异。