Marvelous™ 三通旋塞对血压传感器套件固有频率和阻尼系数的影响。

Influence of the marvelous™ three-way stopcock on the natural frequency and damping coefficient in blood pressure transducer kits.

作者信息

Fujiwara Shigeki Joseph Luke, Tachihara Keiichi, Mori Satoshi, Ouchi Kentaro, Itakura Shoko, Yasuda Michiko, Hitosugi Takashi, Imaizumi Uno, Miki Yoichiro, Toyoguchi Izumi, Yoshida Kazu-Ichi, Yokoyama Takeshi

机构信息

Department of Anesthesiology, Graduate School of Dentistry, Kanagawa Dental University, 82 Inaoka-cho, Yokosuka-shi, Kanagawa, 238-0003, Japan.

Department of Dental Anesthesiology, Faculty of Dental Science, Kyushu University, 3-1-1 Maidashi, Higashi-ku, Fukuoka, 812-8582, Japan.

出版信息

J Clin Monit Comput. 2018 Feb;32(1):63-72. doi: 10.1007/s10877-017-9979-0. Epub 2017 Jan 10.

Abstract

Two types of Planecta™ ports are commonly used as sampling ports in blood pressure transducer kits: a flat-type port (FTP) and a port with a three-way stopcock (PTS). Recently, a new type of three-way stopcock (Marvelous™) has been released as a Planecta™ counterpart, but its effects on the frequency characteristics and reliability of blood pressure monitoring have not been investigated. We assessed the influence of the Marvelous™ stopcock on the frequency characteristics of the pressure transducer kit. The basic pressure transducer kit, DT4812J, was modified by replacing one or two of the original three-way stopcocks with Marvelous™ stopcocks. The frequency characteristics (i.e., natural frequency and damping coefficient) of each kit were determined using wave parameter analysis software, and subsequently evaluated on a Gardner chart. Replacement of the original blood pressure transducer kit stopcocks with Marvelous™ stopcocks decreased the natural frequency (48.3 Hz) to 46.3 Hz or 44.8 Hz, respectively; the damping coefficient was not significantly changed. Plotting the data on a Gardner chart revealed that the changes fell within the adequate dynamic response region, indicating they were within the allowable range. Insertion of Marvelous™ stopcocks slightly affects the natural frequency of the pressure transducer kit, similar to inserting a PTS. The results indicate that the Marvelous™ stopcock is useful for accurate monitoring of arterial blood pressure, and may be recommended when insertion of two or more closed-loop blood sampling systems is necessary.

摘要

在血压传感器套件中,通常使用两种类型的Planecta™端口作为采样端口:扁平型端口(FTP)和带有三通旋塞的端口(PTS)。最近,一种新型三通旋塞(Marvelous™)作为Planecta™的对应产品发布,但尚未研究其对血压监测频率特性和可靠性的影响。我们评估了Marvelous™旋塞对压力传感器套件频率特性的影响。通过用Marvelous™旋塞替换一个或两个原始三通旋塞,对基本压力传感器套件DT4812J进行了改进。使用波形参数分析软件确定每个套件的频率特性(即固有频率和阻尼系数),随后在加德纳图上进行评估。用Marvelous™旋塞替换原始血压传感器套件的旋塞后,固有频率分别从48.3Hz降至46.3Hz或44.8Hz;阻尼系数没有显著变化。将数据绘制在加德纳图上显示,这些变化落在适当的动态响应区域内,表明它们在允许范围内。插入Marvelous™旋塞对压力传感器套件的固有频率影响较小,类似于插入PTS。结果表明,Marvelous™旋塞有助于准确监测动脉血压,在需要插入两个或更多闭环采血系统时可能值得推荐。