Fujiwara Shigeki Joseph Luke, Tachihara Keiichi, Mori Satoshi, Ouchi Kentaro, Itakura Shoko, Yasuda Michiko, Hitosugi Takashi, Imaizumi Uno, Miki Yoichiro, Toyoguchi Izumi, Yoshida Kazu-Ichi, Yokoyama Takeshi
Department of Anesthesiology, Graduate School of Dentistry, Kanagawa Dental University, 82 Inaoka-cho, Yokosuka-shi, Kanagawa, 238-0003, Japan.
Department of Dental Anesthesiology, Faculty of Dental Science, Kyushu University, 3-1-1 Maidashi, Higashi-ku, Fukuoka, 812-8582, Japan.
J Clin Monit Comput. 2018 Feb;32(1):63-72. doi: 10.1007/s10877-017-9979-0. Epub 2017 Jan 10.
Two types of Planecta™ ports are commonly used as sampling ports in blood pressure transducer kits: a flat-type port (FTP) and a port with a three-way stopcock (PTS). Recently, a new type of three-way stopcock (Marvelous™) has been released as a Planecta™ counterpart, but its effects on the frequency characteristics and reliability of blood pressure monitoring have not been investigated. We assessed the influence of the Marvelous™ stopcock on the frequency characteristics of the pressure transducer kit. The basic pressure transducer kit, DT4812J, was modified by replacing one or two of the original three-way stopcocks with Marvelous™ stopcocks. The frequency characteristics (i.e., natural frequency and damping coefficient) of each kit were determined using wave parameter analysis software, and subsequently evaluated on a Gardner chart. Replacement of the original blood pressure transducer kit stopcocks with Marvelous™ stopcocks decreased the natural frequency (48.3 Hz) to 46.3 Hz or 44.8 Hz, respectively; the damping coefficient was not significantly changed. Plotting the data on a Gardner chart revealed that the changes fell within the adequate dynamic response region, indicating they were within the allowable range. Insertion of Marvelous™ stopcocks slightly affects the natural frequency of the pressure transducer kit, similar to inserting a PTS. The results indicate that the Marvelous™ stopcock is useful for accurate monitoring of arterial blood pressure, and may be recommended when insertion of two or more closed-loop blood sampling systems is necessary.
在血压传感器套件中,通常使用两种类型的Planecta™端口作为采样端口:扁平型端口(FTP)和带有三通旋塞的端口(PTS)。最近,一种新型三通旋塞(Marvelous™)作为Planecta™的对应产品发布,但尚未研究其对血压监测频率特性和可靠性的影响。我们评估了Marvelous™旋塞对压力传感器套件频率特性的影响。通过用Marvelous™旋塞替换一个或两个原始三通旋塞,对基本压力传感器套件DT4812J进行了改进。使用波形参数分析软件确定每个套件的频率特性(即固有频率和阻尼系数),随后在加德纳图上进行评估。用Marvelous™旋塞替换原始血压传感器套件的旋塞后,固有频率分别从48.3Hz降至46.3Hz或44.8Hz;阻尼系数没有显著变化。将数据绘制在加德纳图上显示,这些变化落在适当的动态响应区域内,表明它们在允许范围内。插入Marvelous™旋塞对压力传感器套件的固有频率影响较小,类似于插入PTS。结果表明,Marvelous™旋塞有助于准确监测动脉血压,在需要插入两个或更多闭环采血系统时可能值得推荐。