Parati G, Pomidossi G, Casadei R, Mancia G
Hypertension. 1985 Jul-Aug;7(4):597-601. doi: 10.1161/01.hyp.7.4.597.
Intermittent noninvasive blood pressure monitoring is becoming increasingly popular because of the belief that a daytime blood pressure profile can provide a better clinical evaluation of hypertension than that provided by casual blood pressure measurements. This approach has potential limitations, however, one of which is that the cuff inflations permitting blood pressure to be repeatedly measured may induce an alerting reaction and a pressor response in the patients and lead to an overestimation of their daytime blood pressure. Blood pressure in 22 subjects was invasively recorded for 24 hours by the Oxford method. During the day of the recording blood pressure was also measured by a noninvasive device (Vita-Stat 901), which had its cuff applied to the opposite arm from which the intra-arterial signal was derived. For 2 hours the device provided automatic cuff inflations at 10-minutes intervals. For another 2 hours it was programmed to provide cuff inflations only following patients' commands, also at 10-minute intervals. Analysis of the intra-arterial blood pressure trace during the periods preceding and following the automatically or semiautomatically induced cuff inflations showed that these procedures caused no increment in systolic and diastolic blood pressure. This finding applied not only to the mean data but also to each individual measurement considered separately, including the initial one. Our results indicate that automatic and semiautomatic blood pressure monitorings do not induce an alarm reaction and a blood pressure rise and thus do not overestimate daytime blood pressure values.
间歇性无创血压监测正变得越来越流行,因为人们相信与偶尔测量的血压相比,日间血压概况能对高血压提供更好的临床评估。然而,这种方法存在潜在局限性,其中之一是允许反复测量血压的袖带充气可能会在患者中引发警觉反应和升压反应,并导致对其日间血压的高估。采用牛津方法对22名受试者的血压进行了24小时有创记录。在记录当天,还通过无创设备(Vita-Stat 901)测量血压,该设备的袖带应用于与获取动脉内信号的手臂相对的另一侧手臂。该设备在两小时内每隔10分钟自动进行一次袖带充气。在另外两小时内,它被设定为仅在患者指令后进行袖带充气,同样是每隔10分钟一次。对自动或半自动诱导袖带充气之前和之后的动脉内血压轨迹分析表明,这些操作不会导致收缩压和舒张压升高。这一发现不仅适用于平均数据,也适用于单独考虑的每个个体测量值,包括初始测量值。我们的结果表明,自动和半自动血压监测不会引发警报反应和血压升高,因此不会高估日间血压值。