Tjörnstrand B, Lidgren L
Acta Orthop Scand. 1985 Apr;56(2):124-6. doi: 10.3109/17453678508994334.
Three cases of total knee replacement, in which the polyethylene tibial component was fractured, are presented. Two were Attenborough prostheses and the third was a Townley prosthesis. In all cases there were radiographic signs of loosening before the fracture occurred. They were all successfully reoperated with an arthrodesis, using the long Küntscher nail technique.