Temporal bone image quality in CBCT: Device and protocol variations.

作者信息

Razavi Seyed Hossein, Poormohammadi Mohammad, Ansarilari Arezoo

机构信息

Department of Oral and Maxillofacial Radiology, School of Dentistry, Shahid Sadoughi University of Medical Sciences, Yazd, Iran.

出版信息

Cochlear Implants Int. 2025 Jan;26(1):73-78. doi: 10.1080/14670100.2025.2478739. Epub 2025 Mar 29.

Abstract

OBJECTIVES

Considering the high prevalence of temporal bone imaging and the growing tendency to replace the CT scan with CBCT in this field, it seems necessary to determine the optimal radiation conditions in each commercial CBCT brand to obtain the best diagnostic images with minimal patient radiation.

METHODS

A dry human skull was imaged by 6 radiation protocols in 4 different CBCT devices. The images of anatomical landmarks including cochlea, lamina spiralis, facial canal, semicircular canals, and modiolus, were reconstructed. Then, the quality of these images was determined by 3 observers based on a 4-point scoring system. The Kappa coefficient was adopted to evaluate the observer's agreement, and the Kruskal-Wallis and Mann-Whitney tests were used to compare the mean scores of the protocols and the superiority of the protocols in each device, respectively.

RESULTS

There was no statistically significant difference between the mean scores of the protocols in the Acteon and NewTom devices. In the Kodak and Planmeca devices, the recommended protocol for imaging the temporal bone is protocol 1 and protocol 3, respectively.

CONCLUSION

Considering the replacement of the CBCT device for temporal bone imaging and the variety of related protocols in these devices, to reduce the overall society dose, it seems necessary to conduct further studies to determine the optimal radiation conditions in each device. In addition, the ability to unilaterally image the temporal bone is an underappreciated advantage of the CBCT over the CT, which prescribers are not familiar with.

摘要

目的

鉴于颞骨成像的高普及率以及该领域中用锥形束计算机断层扫描(CBCT)取代CT扫描的趋势日益增长,确定每个商用CBCT品牌的最佳辐射条件以获得最佳诊断图像并使患者辐射剂量最小化似乎很有必要。

方法

使用4种不同的CBCT设备,通过6种辐射方案对一个干燥的人类头骨进行成像。重建包括耳蜗、螺旋板、面神经管、半规管和蜗轴在内的解剖标志的图像。然后,由3名观察者根据4分评分系统确定这些图像的质量。采用Kappa系数评估观察者间的一致性,分别使用Kruskal-Wallis检验和Mann-Whitney检验比较各方案的平均得分以及各设备中方案的优越性。

结果

Acteon和NewTom设备中各方案的平均得分之间无统计学显著差异。在柯达和普兰梅卡设备中,用于颞骨成像的推荐方案分别是方案1和方案3。

结论

考虑到CBCT设备用于颞骨成像以及这些设备中相关方案的多样性,为降低全社会的总体剂量,似乎有必要开展进一步研究以确定每个设备的最佳辐射条件。此外,CBCT相对于CT具有单侧成像颞骨的能力,这是一个未得到充分重视的优势,开处方者对此并不熟悉。