Sachs B P, Abitbol M M, Cho G, Acker D
Obstet Gynecol. 1985 Sep;66(3):434-5.
Presented is a preliminary report of the use of a new type of scalp electrode (patent pending), which, because of its design, may have the advantage of improved safety. It was used on 40 patients in labor. To confirm the quality of the recording, in all cases, a fetal electrocardiogram was also obtained. Application was achieved on the first attempt in 37 cases in two additional cases on the second attempt. There were no reported cases of the electrode being dislodged during labor. The quality of the recording was reported to be good. There were no reported cases of trauma to the fetal scalp nor were there any cases of infection noted.
本文呈现了关于一种新型头皮电极(正在申请专利)使用情况的初步报告,该电极因其设计特点,可能具有提高安全性的优势。它被用于40名分娩期患者。为确认记录质量,所有病例均同时获取了胎儿心电图。37例在首次尝试时成功应用,另外2例在第二次尝试时成功。未报告电极在分娩过程中移位的病例。据报告记录质量良好。未报告胎儿头皮受创伤的病例,也未发现感染病例。