Hofmeyr G J, Nikodem V C, Gülmezoĝlu A M, Bunn A E
Department of Obstetrics and Gynaecology, Coronation Hospital, Johannesburg.
Br J Obstet Gynaecol. 1993 Jul;100(7):649-52. doi: 10.1111/j.1471-0528.1993.tb14232.x.
To develop a nonpenetrating scalp electrode for intrapartum fetal monitoring.
Preliminary observational evaluation of the device.
An urban academic hospital in Johannesburg, South Africa.
Fifteen women in labour.
Application of the electrode for intrapartum monitoring.
Quality of tracings, duration of application, disconnection of the electrode, marking of the scalp.
Application was successful in 13 of 15 women, and high quality tracings obtained in 12.
The design of scalp electrode tested is effective and produces high quality tracings in most cases.
研发一种用于产时胎儿监护的非穿透性头皮电极。
对该设备进行初步观察性评估。
南非约翰内斯堡的一家城市学术医院。
15名分娩期妇女。
将电极用于产时监护。
描记质量、应用持续时间、电极断开情况、头皮标记情况。
15名妇女中有13名应用成功,12名获得高质量描记。
所测试的头皮电极设计有效,在大多数情况下可产生高质量描记。