Evaluation of MRI Safety of Ru-106 Eye Applicators.

作者信息

Tang Michael C Y, Klaassen Lisa, Marinkovic Marina, Vu T H Khanh, Luyten Gregorius P M, Creutzberg Carien L, Ketelaars Martijn, Beenakker Jan-Willem M

机构信息

Department of Ophthalmology, Leiden University Medical Center, Leiden, The Netherlands.

Department of Radiology, Leiden University Medical Center, Leiden, The Netherlands.

出版信息

Ocul Oncol Pathol. 2025 Apr;11(1):13-20. doi: 10.1159/000542712. Epub 2024 Nov 30.

Abstract

INTRODUCTION

Ruthenium-106 brachytherapy is a primary treatment for uveal melanoma (UM), the most common intra-ocular malignancy in adults. This study evaluated the safety of Ru-106 applicators at 3 Tesla (T) MRI and their impact on image quality.

METHODS

Magnetic attraction and eddy currents were tested on a 20-mm-diameter Ru-106 applicator using a nylon string and a porcine eye. Safety criteria were defined by ocular oncologists, comparing magnetic field interactions to the forces exerted on the eye during surgery. Five UM patients were scanned at 3T MRI with the applicator in situ using both conventional anatomical sequences and scans optimised to reduce metal artefacts.

RESULTS

Minimal magnetic interactions were observed. Eddy currents caused slight lagging during fast movements and temporary detachment of the applicator of the porcine eye in conditions that were considered unrealistic for clinical scans. Significant susceptibility artefacts compromised image quality of the affected eye.

CONCLUSION

Patients with Ru-106 applicators can be safely used in 3T MRI with some simple precautions. MR image quality of the eye was poor due to major susceptibility artefacts; however, imaging of extra-ocular anatomy is feasible.

摘要

引言

钌 - 106近距离放射疗法是成人最常见的眼内恶性肿瘤——葡萄膜黑色素瘤(UM)的主要治疗方法。本研究评估了钌 - 106施源器在3特斯拉(T)磁共振成像(MRI)中的安全性及其对图像质量的影响。

方法

使用尼龙绳和猪眼对直径20毫米的钌 - 106施源器进行磁吸引力和涡流测试。眼科肿瘤学家定义了安全标准,将磁场相互作用与手术期间施加在眼睛上的力进行比较。五名UM患者在施源器原位的情况下接受3T MRI扫描,使用传统解剖序列和优化以减少金属伪影的扫描。

结果

观察到最小的磁相互作用。在快速移动过程中,涡流导致轻微滞后,并且在被认为对临床扫描不切实际的条件下,猪眼的施源器出现暂时脱离。明显的磁化率伪影损害了患眼的图像质量。

结论

采取一些简单的预防措施后,钌 - 106施源器的患者可以安全地进行3T MRI检查。由于主要的磁化率伪影,眼部的磁共振图像质量较差;然而,眼外解剖结构的成像可行。