Rehr R B, Malloy C R, Filipchuk N G, Peshock R M
Radiology. 1985 Sep;156(3):717-9. doi: 10.1148/radiology.156.3.4023232.
We assessed the potential of proton magnetic resonance (MR) imaging for accurately measuring left ventricular volumes using 15 latex casts of excised human left ventricles. The casts were submerged in water to stimulate the endocardial left ventricular cavity interface in in vivo imaging conditions. Tomographic image sections perpendicular to the long axis of the cast were obtained, spanning each cast from apex to base. Simpson's rule was used to calculate the cast volumes. Correlation between the actual cast volumes (as measured by the displacement method) and the calculated volumes using MR imaging for the 15 casts was excellent. Our data demonstrate that MR imaging accurately measures cardiac chamber volumes in this in vitro model.
我们使用15个切除的人类左心室乳胶铸型,评估了质子磁共振(MR)成像准确测量左心室容积的潜力。将铸型浸没在水中,以模拟体内成像条件下的左心室内膜腔界面。获取垂直于铸型长轴的断层图像切片,从心尖到心底覆盖每个铸型。采用辛普森法则计算铸型容积。15个铸型的实际铸型容积(通过排水法测量)与使用MR成像计算的容积之间的相关性非常好。我们的数据表明,在这个体外模型中,MR成像能够准确测量心腔容积。