经静脉导线拔除术中起搏器导线的比较耐用性:通过台架试验进行评估。

Comparative durability of pacemaker leads in transvenous lead extraction: An evaluation through bench testing.

作者信息

Morita Junji, Okada Ayako, Kusumoto Fred, Nakamura Kentaro

机构信息

Department of Cardiovascular Medicine, Sapporo Cardiovascular Clinic, Sapporo, Hokkaido, Japan.

Department of Cardiovascular Medicine, Shinshu University School of Medicine, Nagano, Japan.

出版信息

Heart Rhythm O2. 2025 Jan 30;6(4):481-488. doi: 10.1016/j.hroo.2025.01.015. eCollection 2025 Apr.

Abstract

BACKGROUND

Transvenous lead extraction (TLE) is a less invasive alternative to surgical removal, but leads may break during TLE.

OBJECTIVE

The purpose of this study was to compare the tensile strength and behavior under stress of leads from different manufacturers during TLE using different extraction techniques.

METHODS

Different lead types (Tendril STS, INGEVITY, INGEVITY+, Solia S, and CapSure Fix) were subjected to tensile testing using a testing machine. The leads were tested in up to 3 different configurations: terminal removed, terminal retained, and removal using a snare. Lead durability was assessed by increasing force and measuring elongation and the force required for complete disruption.

RESULTS

The tensile strength of leads varied greatly based on the extraction technique and lead type. The INGEVITY leads showed improved durability when secured with a snare, despite weak performance when the terminal was removed or retained. The INGEVITY+ lead exhibited enhanced durability, particularly when removed using a snare. The Tendril STS lead exhibited the least durability when the terminal was removed but had comparable durability to the Solia S and CapSure Fix leads when the terminal was retained. The Solia S and CapSure Fix leads had consistent higher durability, irrespective of whether or not the terminal was removed.

CONCLUSION

This study found wide variability of lead behavior among manufacturers and with different extraction techniques. Terminal retention and snare usage enhanced lead durability, suggesting that these techniques should be considered depending on lead type and for leads with longer implantation periods or if higher extraction forces are anticipated.

摘要

背景

经静脉导线拔除术(TLE)是一种侵入性较小的手术拔除替代方法,但导线在TLE过程中可能会断裂。

目的

本研究的目的是比较使用不同拔除技术进行TLE时,不同制造商导线的拉伸强度和应力下的行为。

方法

使用测试机对不同类型的导线(Tendril STS、INGEVITY、INGEVITY+、Solia S和CapSure Fix)进行拉伸测试。导线在多达3种不同配置下进行测试:去除末端、保留末端以及使用圈套器进行拔除。通过增加力并测量伸长量以及完全破坏所需的力来评估导线的耐久性。

结果

导线的拉伸强度因拔除技术和导线类型而异。INGEVITY导线在使用圈套器固定时显示出更好的耐久性,尽管在去除或保留末端时性能较弱。INGEVITY+导线表现出更高的耐久性,特别是在使用圈套器拔除时。Tendril STS导线在去除末端时耐久性最差,但在保留末端时与Solia S和CapSure Fix导线的耐久性相当。Solia S和CapSure Fix导线无论是否去除末端,都具有始终较高的耐久性。

结论

本研究发现不同制造商以及不同拔除技术之间导线行为存在很大差异。保留末端和使用圈套器可提高导线耐久性,这表明应根据导线类型以及植入时间较长或预计需要更高拔除力的导线来考虑这些技术。