Suppr超能文献

使用PyMedPhys对名义上射束匹配的医用直线加速器的治疗计划进行三维伽马分析。

3D gamma analysis between treatment plans for nominally beam-matched medical linear accelerators using PyMedPhys.

作者信息

Guan Fada, Donahue William, Biggs Simon, Jennings Matthew, Draeger Emily, Chen Huixiao, Wang Yuenan, Nguyen Ngoc, Carlson David J, Chen Zhe, Han Dae Yup

机构信息

Department of Therapeutic Radiology Yale School of Medicine New Haven Connecticut USA.

Department of Radiation Physics MD Anderson Cancer Center Houston Texas USA.

出版信息

Precis Radiat Oncol. 2024 Nov 27;8(4):191-199. doi: 10.1002/pro6.1247. eCollection 2024 Dec.

Abstract

Beam-matched linear accelerators (linacs) enable flexible patient scheduling and efficient treatment delivery in the event of unexpected machine downtime. The purpose of this study was to test the feasibility of 3D gamma index as an additional metric beyond standard measurement-based comparisons for more efficient evaluation of treatment plans between linacs with nominally matched beam models to ensure safe patient transfer. Seventeen 3D conformal radiotherapy (3DCRT) plans and thirty-six volumetric-modulated radiation therapy (VMAT) plans for different disease sites were selected from the original linac. An in-house script was used to automatically create new plans for the target linac and calculate dose using parameters of the original plans. 3D gamma analysis was performed to compare plan dose distributions between the target and original linacs using . The 2%/2 mm gamma pass (γ≤1) rate was >99.99% for all 3DCRT plans. The median 1%/1 mm pass rate was 99.86% but two cases failed (< 90%). For VMAT plans, the median and minimum 2%/2 mm gamma pass rates were 99.43% and 93.81%. For 1%/1 mm, the median pass rate was 92.02% but ten cases failed. The results indicated using 3D gamma index can enhance the confidence and add an extra layer for safe patient transfer.

摘要

束流匹配的直线加速器(直线加速器)在机器意外停机时能够实现灵活的患者调度和高效的治疗交付。本研究的目的是测试三维伽马指数作为标准测量比较之外的附加指标的可行性,以便更有效地评估标称束流模型匹配的直线加速器之间的治疗计划,确保患者安全转移。从原始直线加速器中选择了针对不同疾病部位的17个三维适形放疗(3DCRT)计划和36个容积调强放疗(VMAT)计划。使用内部脚本自动为目标直线加速器创建新计划,并使用原始计划的参数计算剂量。使用进行三维伽马分析,以比较目标直线加速器和原始直线加速器之间的计划剂量分布。所有3DCRT计划的2%/2mm伽马通过率(γ≤1)均>99.99%。1%/1mm通过率的中位数为99.86%,但有两例失败(<90%)。对于VMAT计划,2%/2mm伽马通过率的中位数和最小值分别为99.43%和93.81%。对于1%/1mm,通过率的中位数为92.02%,但有10例失败。结果表明,使用三维伽马指数可以增强信心,并为患者安全转移增加一层保障。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/2f53/11934910/2acaa821602c/PRO6-8-191-g004.jpg

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验