Suppr超能文献

一种基于心电图的无创、可穿戴、无线产时监测工具相对于传统基于超声的CTG的准确性。

Accuracy of a Noninvasive, Wearable, Wireless, ECG-Based, Intrapartum Monitoring Tool Against the Conventional Ultrasound-Based CTG.

作者信息

Damodaran Nair Kanjoor Deepthi, Gilvaz Sareena, Menon Bindu, Singh Priyanka

机构信息

Department of Obstetrics and Gynaecology, Jubilee Mission Medical College and Research Institute, Thrissur, Kerala India.

Public Health Manager, Janitri Innovations Pvt Ltd, Bangalore, India.

出版信息

J Obstet Gynaecol India. 2025 Apr;75(Suppl 1):340-347. doi: 10.1007/s13224-024-01998-2. Epub 2024 Aug 31.

Abstract

BACKGROUND

Most conventional cardiotocographs (CTG) work on the ultrasound principle and are heavily used by doctors to measure fetal heart rate (FHR) and uterine contractions (UCs). Despite their widespread usage, they have limitations with accessibility, portability and ease of use. Electrocardiogram (ECG) which relies on electrical signals gives a far superior signal during labor, yet it is ineffective to measure FHR unless an invasive scalp electrode is inserted. Noninvasive fECGs (ni-fECG) are still subject to constant research and validation. The ni-fECG device is a novel, ECG and electromyography (EMG)-based medical device that measures FHR, UC and maternal heart rate (MHR) continuously. It is portable, wearable and easy to use. This study aimed to assess the accuracy of the ni-fECG device with the conventional ultrasound-based CTG.

METHODS

A prospective observational study was conducted on 282 pregnant women. The inclusion criteria included women ≥ 28 weeks of gestational age with singleton pregnancy and cephalic presentation. The procedure used in both the devices to monitor FHR and UC was explained to the participants. Data analysis was done using Python Jupyter Notebook.

RESULTS

The accuracy for baseline heart rate, FHR and UC was found to be 97.37%, 91.3% and 70.27%, respectively. BMI and maternal age did not have a significant impact on the comparative accuracy.

CONCLUSION

With further improvements in UC accuracy the ni-fECG device could be a viable alternative to CTGs.

摘要

背景

大多数传统的胎心宫缩图(CTG)基于超声原理工作,被医生大量用于测量胎儿心率(FHR)和子宫收缩(UC)。尽管它们被广泛使用,但在可及性、便携性和易用性方面存在局限性。依靠电信号的心电图(ECG)在分娩期间能提供质量高得多的信号,然而,除非插入侵入性头皮电极,否则它无法有效测量胎儿心率。非侵入性胎儿心电图(ni-fECG)仍在不断研究和验证中。ni-fECG设备是一种基于心电图和肌电图(EMG)的新型医疗设备,可连续测量胎儿心率、子宫收缩和母体心率(MHR)。它便于携带、可穿戴且易于使用。本研究旨在评估ni-fECG设备与传统超声CTG的准确性。

方法

对282名孕妇进行了一项前瞻性观察研究。纳入标准包括孕周≥28周的单胎妊娠且头位的妇女。向参与者解释了两种设备用于监测胎儿心率和子宫收缩的程序。使用Python Jupyter Notebook进行数据分析。

结果

发现基线心率、胎儿心率和子宫收缩的准确率分别为97.37%、91.3%和70.27%。体重指数和产妇年龄对比较准确性没有显著影响。

结论

随着子宫收缩准确性的进一步提高,ni-fECG设备可能成为CTG的可行替代方案。

相似文献

3
Discontinuation of intravenous oxytocin in the active phase of induced labour.引产活跃期静脉滴注缩宫素的停用
Cochrane Database Syst Rev. 2018 Aug 20;8(8):CD012274. doi: 10.1002/14651858.CD012274.pub2.
4
Fetal scalp stimulation for assessing fetal well-being during labour.胎儿头皮刺激评估分娩时胎儿的健康状况。
Cochrane Database Syst Rev. 2023 Jan 10;1(1):CD013808. doi: 10.1002/14651858.CD013808.pub2.
8
Nitric oxide donors for cervical ripening and induction of labour.用于宫颈成熟和引产的一氧化氮供体。
Cochrane Database Syst Rev. 2016 Dec 5;12(12):CD006901. doi: 10.1002/14651858.CD006901.pub3.

本文引用的文献

9
Understanding fetal physiology and second line monitoring during labor.了解胎儿生理学及分娩期间的二线监测。
J Gynecol Obstet Hum Reprod. 2017 Feb;46(2):113-117. doi: 10.1016/j.jogoh.2016.11.005. Epub 2017 Jan 30.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验