MOLECUBES β-CUBE-一种利用单片 LYSO 闪烁探测器的高空间分辨率和高灵敏度小动物 PET 扫描仪的性能评估。

Performance evaluation of the MOLECUBES β-CUBE-a high spatial resolution and high sensitivity small animal PET scanner utilizing monolithic LYSO scintillation detectors.

机构信息

Department of Radiology, University of Pennsylvania, Philadelphia, PA, United States of America.

出版信息

Phys Med Biol. 2018 Jul 27;63(15):155013. doi: 10.1088/1361-6560/aacec3.

Abstract

The MOLECUBES β-CUBE scanner is the newest amongst commercially available preclinical PET scanners for dedicated small animal imaging. The scanner is compact, lightweight and utilizes a small footprint to facilitate bench-top imaging. It can be used individually, or in combination with the X-CUBE CT scanner, which provides the ability to perform all necessary PET data corrections and provide fully quantitative PET images. The PET detector comprises of an 8 mm thick monolithic LYSO scintillator read-out by an array of 3 mm  ×  3 mm Hamamatsu silicon photomultipliers. The monolithic scintillator provides the ability to measure depth-of-interaction which aids in the development of such a compact scanner. With a scanner diameter of 7.6 cm and axial length of 13 cm it is suitable for imaging both whole-body mice and rats. This paper presents the design and imaging performance of the β-CUBE scanner. NEMA NU4-2008 characterization and a variety of phantom and animal imaging studies to demonstrate the quantitative imaging performance of the PET scanner are presented. Spatial resolution of 1 mm is measured with a filtered-back projection reconstruction algorithm at the center of the scanner and DOI measurement helps maintain the excellent spatial resolution over the entire imaging FOV. An absolute peak sensitivity of 12.4% is measured with a 255-765 keV energy window. The scanner demonstrates good count-rate performance, with a peak NEC of 300 kcps and 160 kcps measured with ~900 µCi in the NEMA mouse and rat phantoms, respectively. Imaging data with the NEMA image quality phantom and Micro Derenzo phantoms demonstrate the ability to achieve good image quality and accurate quantitative data. Image uniformity of 7.4% and spill-over ratio of 8% were measured. The superior spatial resolution, excellent energy resolution and sensitivity also provide superior contrast recovery, with ~70% recovery for the 2 mm rods. While current commercial preclinical PET scanners have spatial resolution in the 1-2 mm range, the 1 mm volumetric resolution presents significant improvement over current commercially available preclinical PET scanners. In combination with the X-CUBE scanner it provides the ability to perform fully quantitative imaging with spatially co-registered high-resolution 3D PET-CT images.

摘要

MOLECUBES β-CUBE 扫描仪是目前市场上最新的专用小动物成像临床前 PET 扫描仪。该扫描仪结构紧凑、重量轻,占地面积小,便于在台式机上进行成像。它可以单独使用,也可以与 X-CUBE CT 扫描仪结合使用,后者可以提供执行所有必要的 PET 数据校正并提供完全定量的 PET 图像的功能。PET 探测器由 8mm 厚的整体式 LYSO 闪烁体组成,由一个 3mm×3mm 的 Hamamatsu 硅光电倍增管阵列进行读取。整体式闪烁体提供了测量相互作用深度的能力,这有助于开发如此紧凑的扫描仪。该扫描仪的直径为 7.6cm,轴向长度为 13cm,适用于全身成像小鼠和大鼠。本文介绍了 β-CUBE 扫描仪的设计和成像性能。介绍了 NEMA NU4-2008 特性以及各种体模和动物成像研究,以证明 PET 扫描仪的定量成像性能。在扫描仪的中心位置,使用滤波反投影重建算法测量到 1mm 的空间分辨率,而 DOI 测量有助于在整个成像视场中保持优异的空间分辨率。使用 255-765keV 的能量窗口测量到的绝对峰值灵敏度为 12.4%。该扫描仪具有良好的计数率性能,在 NEMA 小鼠和大鼠体模中分别测量到的峰值 NEC 为 300k cps 和 160k cps,使用的放射性活度约为 900µCi。使用 NEMA 图像质量体模和 Micro Derenzo 体模进行的成像数据表明,该扫描仪能够实现良好的图像质量和准确的定量数据。测量到的图像均匀性为 7.4%,溢出比为 8%。优异的空间分辨率、优异的能量分辨率和灵敏度还提供了更好的对比度恢复,对于 2mm 棒的恢复度约为 70%。虽然目前的商用临床前 PET 扫描仪的空间分辨率在 1-2mm 范围内,但 1mm 的体积分辨率与目前商用临床前 PET 扫描仪相比有了显著的提高。与 X-CUBE 扫描仪结合使用,它可以提供具有空间配准的高分辨率 3D PET-CT 图像的完全定量成像功能。

相似文献

引用本文的文献

1
Spatial Multiplexing and Omics.空间复用与组学
Nat Rev Methods Primers. 2024;4(1). doi: 10.1038/s43586-024-00330-6. Epub 2024 Aug 1.

本文引用的文献

9
Performance evaluation of the small-animal nanoScan PET/MRI system.小动物 nanoScan PET/MRI 系统的性能评估。
J Nucl Med. 2013 Oct;54(10):1825-32. doi: 10.2967/jnumed.112.119065. Epub 2013 Aug 29.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验